Abstract
Acceptable and reproducible radiochemical purity (RP) for 68Ga-DOTATATE was difficult to obtain with the NETSPOT kit because the manufacturer instructions lacked details on the heater or needles used. Methods: The drug was prepared in an International Organization for Standardization (ISO) 5 environment. Multiple dry baths and needle types were used to investigate the effects of reaction temperature and metal contamination, respectively. Temperature curves were obtained with a calibrated thermocouple. The influence of the accuracy of the NETSPOT reagent volume and its effect on outcome were investigated. Results: The AccuBlock dry bath required recalibration for the ISO 5 environment; after calibration, the temperature was stable (only ±0.1°C from the set point). When we followed package insert recommendations (dry bath temperature set to 98°C, reaction time of 8 min), the reaction temperature was 90.6°C. When Becton Dickinson needles were used for reconstitution, 15 of 18 runs (83%) did not meet the RP specification. However, B. Braun Medical needles achieved satisfactory and stable RP. When the 68Ga generator was eluted with 5.0 mL of 0.1 M hydrochloric acid (HCl), only 3.8–3.9 mL of eluate reached the reaction vial; this volume did not impact labeling (final pH was 3.8). The labeling success rate increased markedly if the 68Ga eluate was passed through a conditioned silica gel cartridge or if no cartridge was used; then, RP was more than 99%. HCl contact with the septum of the labeling vial reduced RP. Conclusion: The needle type and the temperature setting of the dry bath have critical roles in 68Ga-DOTATATE preparation. The AccuBlock dry bath has excellent stability and accuracy and can be used for reliable preparation. By using a conditioned silica gel cartridge or by eliminating the cartridge altogether, the RP is reliably high and stable.
Neuroendocrine tumors have receptors for somatostatin, a hormone that regulates the endocrine system. A somatostatin receptor imaging agent was approved in 2016 by the U.S. Food and Drug Administration (U.S. FDA) for PET imaging of neuroendocrine tumors in adult and pediatric patients (1). The imaging agent 68Ga-DOTATATE is a radioconjugate consisting of the somatostatin analog tyrosine-3-octreotate labeled with the tracer 68Ga via the macrocyclic chelating agent dodecanetetraacetic acid. In terms of convenience in patient care, 68Ga-DOTATATE represents a major advantage because imaging can be completed in less than 2 h; in comparison, an octreotide scan requires 2 d (2).
In our initial experience preparing 68Ga-DOTATATE following the manufacturer instructions, we encountered numerous problems that resulted in low radiochemical purity (RP) and low rates of successful preparation. Notably, the instructions did not specify a temperature for the reaction solution, nor did they specify manufacturer or model names for the dry bath or sterile needles. To establish a reliable, high-quality production method, we evaluated the effects of 3 reaction conditions. The first was the dry-bath temperature and the actual temperature of the reaction solution, the second was metal contamination from needles, and the third was the accuracy of the NETSPOT reagent volume and the effect of volume on outcome. In this study, we aimed to systematically evaluate factors that affect preparation of 68Ga-DOTATATE and establish guidelines that would increase the rate of successful preparations for new users.
MATERIALS AND METHODS
68Ga-DOTATATE was prepared with a GalliaPharm 68Ge/68Ga generator (Eckert and Ziegler Radiopharma), sterile ultrapure 0.1 M hydrochloric acid (HCl), and the NETSPOT kit (Advanced Accelerator Applications). We followed the NETSPOT manufacturer instructions (3) and prepared 68Ga-DOTATATE in an aseptic International Organization for Standardization (ISO) 5 isolator with laminar flow. Briefly, we used 5.0 mL of 0.1 M HCl to elute 68GaCl3 from the 68Ge/68Ga generator, and for some preparations, the eluate was passed through a silica gel cartridge. The cartridge was connected to a sterile, siliconized needle plus a sterile filter and then connected to the reaction vial of the NETSPOT kit. With a vent filter (Millipore Millex-FG) and needles (B. Braun Medical 100 Sterican [0.60 × 60 mm] or Becton Dickinson Precisionglide hypodermic [siliconized] [0.6 × 60 mm]), this elution procedure took 2.5 min. Within 10 min of adding 68GaCl3 to the reaction vial, 0.5 mL of reaction buffer was injected and the vial was heated in a dry bath (set-point range, 95°C–98°C). The instructions indicated heating for at least 7 min but no longer than 10 min.
RP was determined with thin-layer chromatography. Glass microfiber chromatography paper impregnated with silica gel (Agilent Technologies, Inc.) was used for the stationary phase, and 1 M ammonium acetate:methanol (1:1 v/v) was used for the mobile phase. According to the revised NETSPOT package insert (4), 68Ga DOTATATE RP should be at least 95% and other 68Ga species should be no more than 5%. The success rate was defined as the percentage of production runs that met the minimal RP.
Temperature Control
Temperature curves were obtained using NETSPOT reaction vials containing 5.0 mL of water and dry baths set to 95°C and 98°C. Temperatures were verified with a calibrated thermocouple thermometer (Fluke 54 2 II B). We tested 2 dry heaters: HotPot (Capintec, Inc) and AccuBlock D1200 (Labnet International, Inc.). Plastic covers were used to reduce heat loss and stabilize the temperature in the ISO 5 environment.
68Ge Breakthrough and Metal Content
We measured eluate samples from the 68Ge/68Ga generator weekly to assess 68Ge breakthrough and metal content. Samples were obtained after the generator had not been eluted for more than 48 h, and eluates were not passed through a silica cartridge. The breakthrough analysis was performed by an approved outside vendor on a decayed sample of the drug product. Briefly, the numbers of γ-rays were measured for a specific time as a function of energy (of each γ-ray). A series of calibration procedures then reduced the γ-events to a quantitative measurement of sample activity. Eluates from 2 generators were fully tested with γ-spectrum analysis to verify that 68Ge breakthrough was within specified limits (<0.00100%).
Zinc and iron ions compete with Ga3+ in the labeling reaction and can cause reaction failure. We evaluated the influence of metal contaminants from Becton Dickinson needles and from B. Braun needles. Zinc and iron ion samples (the same as those used for 68Ge breakthrough testing) were analyzed using inductively coupled plasma mass spectrometry.
Cartridge Conditions and Volume and pH of the Final Product
We tested several cartridge conditions: unconditioned silica gel, silica gel conditioned with 0.1 M HCl (2.0 mL), and silica gel conditioned with water (2.0 mL) followed by 5.0 mL of air. We additionally assessed the preparation when no cartridge was used.
We used 5.0 mL of 0.1 M HCl to elute the generator. Depending on the cartridge condition, different volumes of GaCl3 solution were eluted into the reaction vial. The reaction pH was measured with pH test paper (Advatec BPB, pH 2.8–4.4).
Endotoxin Test
Endotoxin testing was conducted using the Endosafe multiple-cartridge system (Charles River).
RESULTS
Temperature Control
The ISO 5 isolator is an environment with a high laminar airflow. Because the airflow is very strong (1,000 air changes per hour), it can affect heat exchange and heat loss with dry baths, particularly at higher temperatures, resulting in temperature deviations from the set point. We assessed the temperature deviation and temperature changes using a NETSPOT reaction vial containing 5.0 mL of water. The AccuBlock dry bath had a highly stable temperature, with only a ±0.1°C deviation from the set point. In contrast, the HotPot showed variations of ±3.0°C from the set point. Given the superior temperature stability of the AccuBlock, the remaining assessments were performed with the AccuBlock only. Routine production also was conducted only with the AccuBlock.
The AccuBlock was factory-calibrated at 40°C. At a 98°C set point, the actual temperature was 94.5°C (i.e., deviated by −3.5°C). After recalibration, the dry bath set at 98°C had deviations of ±0.1°C, and a temperature curve showed that the water in the vial reached the set-point temperature after approximately 21–22 min (Fig. 1).
A representative of the manufacturer indicated that the reaction temperature should range from 90°C to 98°C (pharmacist [operations manager], Advanced Accelerator Applications, oral communication, November 2016); the representative did not indicate an optimal temperature. We set the dry bath to temperatures from 95°C to 105°C and ensured that the set-point temperature was attained before a reaction vial was placed. We then measured how much time was needed for the temperature of the water in the reaction vial to exceed 90°C (range, 6–8 min) (Fig. 2). The water temperature after 8 min ranged from 87.8°C to 97.9°C for set points ranging from 95°C to 105°C (Fig. 2).
For routine preparation of 68Ga-DOTATATE, the dry bath was set to 98°C and the reaction vial was heated for 8 min. We selected these parameters because 98°C was the highest temperature recommended by the manufacturer and the water temperature in the reaction vial was still less than 90°C after a 7-min incubation.
68Ge Breakthrough
The first generator was eluted about 480 times in 375 d, and the 68Ge breakthrough was less than 0.00031% throughout. The second generator was eluted about 720 times in 339 d; the 68Ge breakthrough increased to 0.00060% after 590 elutions (280 d), but it was still within the 0.00100% limit (Fig. 3) (5).
Metal Content
Metal contamination can compete with 68Ga for the chelator group. Thus, the weekly samples also were tested for iron and zinc content (Fig. 4). During the first 2 mo, 68Ga eluates were taken from the first generator with a Becton Dickinson needle and collected in a glass vial with a rubber stopper. However, we realized that glass and rubber could cause high iron and zinc contamination; subsequently, B. Braun needles and metal-free plastic tubes were used for sample collection. We also tested 0.1 M HCl acid passed through a Becton Dickinson needle or a B. Braun needle and 0.1 M HCl solution that was in contact with the rubber septum of a NETSPOT reaction vial (Table 1). The zinc content of acid that had contacted the vial septum was up to 598-fold higher than that of the normal eluate (4.83 vs. 2,887.13 ng/mL); iron content also was 11-fold higher (6.00 vs. 66.23 ng/mL).
Cartridge Conditions
For 33 preparations, the silica cartridge was conditioned with 2.0 mL of 0.1 M HCl and yielded a mean RP of 99.58% (SD, 0.27%) and a success rate of 100%. For 271 preparations, cartridges were flushed with 2 mL of sterile water and dried with 5 mL of air; these preparations yielded a mean RP of 99.08% (SD, 2.58%) and a success rate of 97.79%. We noted only 6 failures in the preparation of 68Ga-DOTATATE that used conditioned cartridges; of these, 4 were operator-related failures. The mean RP for all conditioned cartridges (HCl and water) was 99.13%, and the success rate was 98.03% (298/304). For the 476 preparations using dry cartridges (without conditioning), the mean RP was 97.18% (SD, 5.42%) (this RP was ∼2.0% lower than that for conditioned cartridges), and the success rate was 93.07% (Table 2). We suspect that an unconditioned cartridge may cause a labeling problem.
The 2019 version of the manufacturer instructions indicated that 68Ga eluates could be obtained without using a silica gel cartridge (4). Weekly analysis of 68Ge breakthrough showed levels well below the specific value of 0.00100%. Without the cartridge, the labeling process was more reliable. The setup procedure also was shortened (fewer steps), and elimination of the cartridge when eluting from the 68Ga generator further improved the RP (mean, 99.37% [SD, 1.45%]) and success rate (99.11%) compared with conditioned cartridges. Furthermore, without the cartridge, the final production volume was 5.5 mL, and the pH was within the acceptable range. The increased success rate also improved patient service (e.g., fewer treatment delays, greater patient volume).
Volume and pH of Final Product
The manufacturer instructions did not indicate the optimal final product volume. We used a syringe without natural rubber latex to measure 5.0 mL of 0.1 M HCl and applied that to elute the 68Ga generator. Of the initial 5.0-mL volume, only 3.8–3.9 mL of 68GaCl3 reached the reaction vial because a porous silica cartridge was used to trap the breakthrough 68Ge and a filter was used to sterilize the eluent. Thus, the silica cartridge retained about 1.1 mL of eluent. We added 0.5 mL of buffer to the reaction vial before heating, which made the final volume of the 68Ga-DOTATATE injection approximately 4.4 mL and the pH 3.6–3.8. If a conditioned silica gel cartridge was used, only 4.5–4.6 mL of eluent reached the reaction vial; with 0.5 mL of buffer, the final volume was approximately 5.1 mL and the pH 3.6. If no cartridge was used, then all 5.0 mL of eluent were in the vial; with 0.5 mL of buffer, the final volume was 5.5 mL and the pH 3.4. The acceptable pH range for the reaction is 3.2–3.8. The differences in final volume affected only the pH and did not influence the labeling yield or RP.
Figure 5 shows the RPs of different lots of NETSPOT kits. The mean RP was 97.18%, and the success rate was 94.40% for the 250 preparations before lot PG1917017 was put into use. For lot PG1917017, the mean RP was 99.11% and the success rate was 96.43% (54/56). After the next lot, PG1917023, was put into use. Before conditioned cartridges were used in the reconstruction, the mean RP was 96.54% and the success rate was 95.03% (153/161). When conditioned cartridges were used, the RP was 99.40% and the success rate was 99.33% (298/300). When 68Ga-DOTATATE was reconstituted without a cartridge, the average RP was 99.37% and the success rate of 99.11% (222/224) was the highest among all tested conditions.
Endotoxin Test
The 68Ga-DOTATATE solution inhibited the endotoxin-spike sample test when the solution was diluted 1:100. However, when the sample was diluted 1:400, the solution did not show inhibition.
DISCUSSION
The manufacturer-recommended reaction temperature for 68Ga-DOTATATE labeling should be between 90°C and 98°C. If the temperature of the heater and the heating time of the reaction vial are maintained within the appropriate range, the success rate for production will be higher (more reliable). Because different working environments will affect the temperature differently, we recommend recalibrating the heater in the appropriate environment. For heater set points of 95°C and 98°C, the reaction temperatures after incubating for 8 min were 87.7°C and 90.6°C, respectively. For routine preparations, we used a 98°C set point and heated the reaction vial for 8 min.
To assess the impact of metals, we analyzed only the iron and zinc levels from 68Ge/68Ga generator eluates obtained with Becton Dickinson or B. Braun needles. Iron levels were more than 50-fold higher with Becton Dickinson needles than with B. Braun needles, but zinc levels were similar for both needle types. When Becton Dickinson needles were used, the labeling success rate was only 16.7% (15/18 production runs did not meet the RP specification). However, B. Braun needles consistently achieved sufficient RP. Although both needle types are made of the same type of stainless steel (304 grade) and protected by a siliconized layer, the B. Braun needles had superior performance. We suspect that the internal and external surfaces of the Becton Dickinson needle are not fully siliconized and that, consequently, metals may be leaching from the unprotected surfaces and competitively interfering with 68Ga labeling. We also assessed a 0.1 M HCl solution that was in contact with the rubber septum of the NETSPOT reaction vial and noted high iron and zinc levels. We believe that HCl caused zinc stearate (a common constituent of rubberized septa) to leach into the reaction solution. It is possible that other metal contaminants were present that could affect labeling, but their levels were not measured. We note that Cu2+, if present, also may be competing with Ga3+ at 95°C (6).
One specific lot of NETSPOT kits (lot PG1917017, obtained August 2017) had a notably high and consistent RP (>99.6%). We asked a manufacturer representative whether any changes had been made to the kits, and the representative indicated that the only change was a new lot of silica gel cartridges. The production had an even higher yield when the cartridge was conditioned with 0.1 M HCl or water. However, the effect of the silica gel cartridge on the labeling reaction remains unclear. In August 2018, the U.S. FDA approved a change in how 68Ga-DOTATATE was synthesized with the NETSPOT kit, by allowing removal of the silica gel cartridge during 68Ga generator elution. The average RP without a cartridge exceeded 99.0%. The weekly elution without silica gel cartridges (Fig. 3, generator 2) was closely monitored for 68Ge breakthrough. After 590 elutions, 68Ge breakthrough had increased to up to two thirds of the allowable limit; in that setting, the silica cartridge does have a role as a safety net. We acknowledge that the generator manufacturer recommends no more than 400 elutions per generator (with 5.0 mL of 0.1 M HCl per elution). However, because of a global shortage of 68Ga generators, we eluted the generator more than 400 times and monitored 68Ge breakthrough.
CONCLUSION
The B. Braun needle and the reaction temperature have critical roles in 68Ga-DOTATATE preparation. The internal and external surfaces of the Becton Dickinson needle likely are not fully siliconized, and we surmise that metals leaching from the unprotected surfaces competitively interfere with 68Ga labeling. During our investigations, we also discovered that contact of 0.1 M HCl with the septum of the reaction vial also reduces RP, likely because zinc stearate is leached from the rubberized septa by HCl. The AccuBlock dry bath, plus a cover, has superior stability and an accurate temperature when placed in a high–air-flow isolator. Elimination of the silica cartridge or use of a conditioned cartridge when preparing 68Ga-DOTATATE can markedly increase the RP and production reliability.
“Do” list:
Do use the AccuBlock or Labnet dry bath.
Do recalibrate the dry bath at the high temperature setting and in a high–air-flow environment.
Do use the B. Braun 100 Sterican 0.60 × 60 mm 23-gauge × 2⅜-in needles for the preparation.
“Do-not-do” list:
Do not allow the liquid in the reaction vial to contact the septa before it is heated.
Do not use the Becton Dickinson needles for the preparation.
DISCLOSURE
Advanced Accelerator Applications supplied research-use NETSPOT kits. Mayo Clinic does not endorse the products mentioned in this article. No other potential conflict of interest relevant to this article was reported.
Acknowledgments
We thank Advanced Accelerator Applications for providing expert advice. Portions of this article were published in abstract form (J Nucl Med. 2017;58[suppl 1]:684) and presented at the annual meeting of the Society of Nuclear Medicine and Molecular Imaging, Denver, Colorado, June 10–14, 2017.