Journal of Nuclear Medicine Technology
  • SNMMI
    • JNMT
    • JNM
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • My Cart
Journal of Nuclear Medicine Technology

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • Continuing Education
    • JNMT Podcast
    • SNMMI Annual Meeting Abstracts
  • Subscriptions
    • Subscribers
    • Rates
    • Journal Claims
    • Institutional and Non-member
  • Authors
    • Submit to JNMT
    • Information for Authors
    • Assignment of Copyright
    • AQARA Requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
    • Corporate & Special Sales
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • Watch or Listen to JNMT Podcast
  • Visit SNMMI on Facebook
  • Join SNMMI on LinkedIn
  • Follow SNMMI on Twitter
  • Subscribe to JNMT RSS feeds
Research ArticleIMAGING

Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging

Hanna Piwowarska-Bilska, Leszek J. Hahn, Bozena Birkenfeld, Katarzyna Cichon-Bankowska, Maria H. Listewnik and Piotr Zorga
Journal of Nuclear Medicine Technology December 2010, 38 (4) 181-185; DOI: https://doi.org/10.2967/jnmt.109.073486
Hanna Piwowarska-Bilska
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Leszek J. Hahn
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Bozena Birkenfeld
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Katarzyna Cichon-Bankowska
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Maria H. Listewnik
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Piotr Zorga
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

Abstract

This study was performed to find the optimal low-dose CT protocol for children being imaged on SPECT/CT scanners not equipped with automatic dose control. For SPECT/CT systems with manually adjustable x-ray tube voltage (kV) and anode current (mA), an optimized protocol makes it possible to minimize the dose to patients. Methods: Using the 4-slice low-dose CT component of a commercially available SPECT/CT scanner, we compared the signals reaching the CT detector after radiation passes through objects of different sizes. First, the exit dose rates were measured for combinations of available voltages and currents. Next, imaging parameters were selected on the basis of acceptable levels of exit dose rates, cylindric phantoms of different diameters approximating children of different sizes were scanned using these parameters, and the quality of the CT images was evaluated. Finally, weighted CT dose indexes for abdomen and head CT dose phantoms simulating, respectively, adult and pediatric patients were measured using exactly the same techniques to estimate and compare doses to these 2 groups of patients. Results: For children with torsos smaller than 150 mm, imaging can be performed using the lowest available voltage and current (120 kV and 1 mA, respectively). For children with torsos less than 250 mm, 140 kV and 1.5 mA can be used. For patients with torsos greater than 250 and less than 300 mm, 140 kV and 2 mA can be used. Regarding the signal-to-noise ratio, all these parameters give an excellent signal and fully acceptable noise levels. Conclusion: For the SPECT/CT system studied, even the lowest available voltage and current used for scanning pediatric patients did not cause signal-to-noise degradation, and the use of these settings substantially lowered the dose to the patients.

  • low-dose CT
  • SPECT/CT
  • pediatric radiation exposure
  • CT dose optimization

Diagnostic SPECT/CT systems allow CT scans to be registered with nuclear medicine scans so that radioisotope uptake can be localized. CT scans are additionally used for attenuation correction of radioisotope images. However, compared with SPECT-only examinations, SPECT/CT examinations increase the radiation exposure of patients. SPECT/CT systems with adjustable x-ray tube voltage (kV) and anode current (mA) make it possible to minimize the dose to patients. Reasonable reduction of CT-related dose in individual patients is an important issue, particularly in children (1,2). The aim of this study was to find the optimal pediatric CT protocol that reduces radiation risk.

MATERIALS AND METHODS

All investigations used the 4-slice low-dose CT component of the Infinia Hawkeye 4 SPECT/CT system (GE Healthcare). In this system, both the voltage and the current are adjustable. To compare the signals reaching the CT detector after radiation passes through objects of different sizes, we measured the exit dose for all combinations of available voltages and currents. To simulate patients of different sizes, we used phantoms containing various volumes of water. During all measurements, the x-ray tube was immobilized at the 12-o'clock position in the scanner service mode. To validate this approach and to find the relationship between exit dose rates (detector signal) and CT image quality, we scanned 4 cylindric phantoms filled with water. For this purpose, the system was switched from stationary to tomographic mode, and noise levels representing signal-to-noise ratios in CT images of these phantoms were measured. The diameters of these phantoms corresponded to children of different weights and sizes. To estimate the doses to large and small patients imaged at variable voltages and currents, we scanned the abdomen and head phantoms and measured the weighted CT dose index (CTDIw).

Measurement of Exit Dose

Radiation exiting the water phantom and reaching the CT detector was measured using a Dosemeter/Ratemeter 2026C (Radcal Corp.) ionization chamber, 20X6-3. For this purpose, 8 combinations of available voltage (120 or 140 kV) and current (1, 1.5, 2, or 2.5 mA) were used at a constant time of 23 s, which corresponds to the time of full gantry rotation in the CT mode of scanning. To estimate the strength of the signal reaching the CT detectors, we calculated dose rates by dividing the measured cumulative dose by that time. An acrylic water-filled cylinder was used as a phantom simulating attenuation by the human body. The cylinder was 330 mm in height and 210 mm in diameter. The cylinder was placed on the patient table, vertically (Fig. 1), with its center coinciding with the geometric center of the x-ray fanbeam. The ionization chamber was positioned directly under the cylinder at its center. The water level in the cylinder was increased in steps of 50 mm, from 0 mm to a maximum of 300 mm. All these measurements were performed twice to check repeatability. Overall, 168 measurements of exit dose were performed under the unchanged geometry.

FIGURE 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 1.

Schematic illustration of exit dose measurement geometry.

Noise Level Evaluation

Noise levels for calculation of signal-to-noise ratios (3) were evaluated by imaging sets of water-filled cylinders 270, 200, 160, and 100 mm in diameter in SPECT/CT mode using a bone SPECT protocol. Voltage and current were identical to those used for the exit dose measurements. The reference to which the data obtained from all scanned cylinders were compared corresponded to an adult patient, as represented by a Data Spectrum torso anthropomorphic phantom. The water-filled part of this phantom was scanned at 140 kV and 2.5 mA, and the time of rotation was 23 s. It was assumed that these parameters were tested in the factory and then selected as optimal for adult patient imaging. The slice thickness and collimation are not adjustable for this SPECT/CT system and for all scans were 5 and 20 mm, respectively. For data analysis, circular regions of interest were used. One large region of interest covering about 80% of the image of a scanned cylinder was used to calculate the average SD in the image. This SD represented the noise level for a particular set of parameters and cylinder scan. To evaluate the range of SDs in different parts of the CT image, the set of smaller regions of interest with a diameter equal to about 40 mm was placed circumferentially (12-, 3-, 6-, and 9-o'clock positions) and along the x- and y-axes of CT images of cylinders and torso phantom. SDs in selected regions of interest for a particular slice were calculated using the system software and recorded from a region-of-interest statistics database.

Measurements of CTDIw

The most commonly used dose descriptor is CT dose index. The CTDIw, measured at the center and periphery of CT dose index phantoms, provides a convenient estimate of patient dose for a procedure (4,5). CTDIw is defined as follows:CTDIW=(1/3)(CTDI100)center+(2/3)(CTDI100)periphery.Eq. 1

CTDI100 is the integral of the dose profile along a line perpendicular to the tomographic plane divided by the product of the nominal tomographic section thickness and the number of tomograms produced in a single axial scan. Because the ionization chamber measures an integrated exposure along its 100-mm length, CTDI100 can be calculated according to the following formula:CTDI100=(f⋅E)(L/T).Eq. 2

f is the conversion factor from exposure (mR) to a dose in air (mGy), E is the measured value of exposure, L is the active length of the pencil ionization chamber, and T is the nominal width of each acquired section.

For CT dose index measurements, a CT dose probe was used: a 10X5-3CT pencil-type ionization chamber with a sensitive length of 100 mm (Radcal). Two phantoms were used: solid polymethylmethacrylate disks measuring 160 mm in diameter (head phantom simulating a child's body) and 320 mm in diameter (abdomen phantom simulating an adult's body), both 140 mm thick. There are 5 holes in each of the phantoms, 4 circumferential (4 mm from the edge) and 1 central, with acrylic rods to plug the holes when not in use. The holes are 13.1 mm in diameter to accommodate the size of a standard CT probe. Because the main aim of this work was to evaluate the dose reduction when small patients (children) are scanned using the lowest available current and voltage, only 2 sets of scanning parameters were tested. First, 140 kV and 2.5 mA (maximal available values) were used for scanning the abdomen and head phantoms. These parameters would correspond to the use of a nonadjustable scanner, independent of a patient's size. Then, the lowest available voltage (120 kV) and current (1 mA) were used for scanning the head phantom only, which would correspond to a scanner that is adjustable according to the weight or size of smaller patients.

RESULTS

Table 1 shows the exit dose rates. These calculations used the arithmetic average of 3 cumulative exit doses measured with the 20X6-3 ionization chamber. Because the dispersion of measured cumulative exit doses was small (coefficient of variation < 1%), it is not shown in Table 1. Figure 2 is the graphic representation of the results in Table 1. The shaded area represents parameters giving a detector signal equal to or higher than that produced using the maximum voltage and current available. Table 2 shows noise levels in CT images of cylindric phantoms of different diameters. They represent children of varying sizes scanned using different parameters. These measurements helped to justify the validity of using the exit dose values as a basis for selecting parameters adequate for a child's size.

FIGURE 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 2.

Semilogarithmic graph of exit dose rates for various voltages and currents vs. increasing depth of water in phantom simulating attenuation in patient.

View this table:
  • View inline
  • View popup
TABLE 1

Average Exit Dose Rates

View this table:
  • View inline
  • View popup
TABLE 2

Signal-to-Noise Results for CT Images of Water Phantoms

CTDIw results for both CT dose phantoms are shown in Table 3.

View this table:
  • View inline
  • View popup
TABLE 3

CTDIw for Various Scanning Parameters and Phantoms

Table 4 contains CT acquisition protocols proposed for technologists performing patient studies on the Infinia Hawkeye 4. The information on weight, age, and corresponding patient dimensions is based on the available literature (6). In this study, the arithmetic mean of the chest and abdomen dimensions was used as the average patient size because all measurements of CT dose index and noise were performed on cylindric phantoms. The relationship between body size and weight for adolescents was extrapolated from the data on children and adult patients because of a lack of relevant data in the literature. The results from Table 1 and information about the relationship between children's ages, weights, and sizes were the basis for the proposed low-dose CT acquisition protocols.

View this table:
  • View inline
  • View popup
TABLE 4

Proposed CT Protocol for SPECT/CT in Children

DISCUSSION

The results of this study indicate the possibility of substantially decreasing doses to patients without losing the detector signal. Thus, there is a chance that the risk in children imaged with SPECT/CT systems can be reduced. Table 1 shows that for water layers thinner than 200 mm, the exit dose measured at the minimum available voltage and current (120 kV, 1 mA, respectively) is stronger than that detected when a 300-mm-thick layer of water is penetrated by radiation generated using maximum voltage and current (140 kV and 2.5 mA, respectively). Drawing a line parallel to the “water level in cylinder” axis to intersect the “object exit dose” axis at a point corresponding to the detected exit dose when using 140 kV and 2.5 mA for objects 300 mm in diameter (which corresponds to penetration of a 300-mm-thick water layer), one can find all the parameters for all object diameters that give a signal equal to or greater than this first one (dashed area in Fig. 2). On this basis, parameters optimizing doses to patients of different sizes can be selected. Thus, for infants 150 mm in diameter or smaller, the technique utilizing the lowest available voltage and current can be used because it still gives an excellent detector signal. For patients greater than 150 mm and up to 200 mm in diameter, 120 kV and 2 mA are sufficient. Patients greater than 200 mm and up to 250 mm can be scanned at 140 kV and 1.5 mA. For patients from 250 to 300 mm (adolescents), scanning at 140 kV and 2 mA is suitable. The fact that these parameters translate into a substantial reduction in doses to children and smaller patients is an important message to users of SPECT/CT systems with adjustable settings for voltage and current.

The signal-to-noise results for scanning various sizes of water-filled phantoms with various parameters (Table 2) supports the exit-dose-level criterion as a basis for selecting the imaging parameters. The noise level in the reference image (torso phantom) averaged 42 and ranged from 35 (minimum, the best) to 52 (maximum, the worst). This level is comparable to results obtained for the phantom with a diameter of 270 mm (the largest) scanned at 140 kV and 1 mA: the noise level in that case averaged 43 and ranged from 36 to 54. The noise levels for all remaining phantoms scanned, even at 120 kV and 1 mA, were better than the reference level. Additionally, comparison with the data from Table 1 and Figure 2 shows that using the exit dose rate as a basis for selection of scanning parameters is actually more restrictive than would be selection based on analysis of noise levels only.

The dose reduction can clearly be seen from the CTDIw results for the various scanning parameters and for the 2 phantoms (Table 3). When one switches from 140 kV and 2.5 mA to 120 kV and 1 mA for children 150 mm in diameter, the dose decreases from 7.23 mGy to 2.01 mGy. For infants smaller than 150 mm, the dose reduction would be even greater.

The results presented in this article should be instructive for producers of cheaper SPECT/CT systems in which the CT component is not equipped with automatic dose control. There is definitely some room for improvement. For imaging of small children, the mAs—as a combination of changes in kV, mA, and times of data acquisition—can be decreased further. The CTDIw is 3.6-fold higher when the parameters are not adjustable than when the parameters can be properly adjusted to the weight and height of the patient. Table 4 presents our proposed protocol for the CT part of the acquisition on an Infinia Hawkeye 4. For clinical acquisitions, technologists can set the optimal current and voltage based on patient size and weight.

CONCLUSION

The radiation dose to children imaged with low-dose CT as part of SPECT/CT studies can be reduced. Proper reduction of the voltage and current of the Infinia Hawkeye 4 for patients of smaller size and weight does not decrease the signal strength reaching the CT detectors and does not affect image quality.

REFERENCES

  1. 1.↵
    Committee to Assess Health Risks from Exposure to Low Levels of Ionizing Radiation, National Research Council of the National Academies. Health Risks from Exposure to Low Levels of Ionizing Radiation: BEIR VII, Phase 2. Washington, DC: National Academies Press; 2006.
  2. 2.↵
    1. Brenner DJ,
    2. Hall E
    . Computed tomography: an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–2284.
    OpenUrlCrossRefPubMed
  3. 3.↵
    1. Sprawls P
    . CT image detail and noise. Radiographics. 1992;12:1041–1046.
    OpenUrlPubMed
  4. 4.↵
    1. Bongartz G,
    2. Golding SJ,
    3. Jurik AG,
    4. et al
    . European Guidelines for Quality Criteria for Computed Tomography: EUR 16262. Available at: http://www.drs.dk/guidelines/ct/quality/. Accessed August 30, 2010.
  5. 5.↵
    1. Shope TB,
    2. Gagne RM,
    3. Johnson GC
    . A method for describing the doses delivered by transmission x-ray tomography. Med Phys. 1981;8:488–495.
    OpenUrlCrossRefPubMed
  6. 6.↵
    1. Kereiakes GJ,
    2. Rosenstein M
    . Handbook of Radiation Doses in Nuclear Medicine and Diagnostic X-Ray. Boca Raton, FL: CRC Press; 1980:187, 214.
    OpenUrl
  • Received for publication December 30, 2009.
  • Accepted for publication July 2, 2010.
PreviousNext
Back to top

In this issue

Journal of Nuclear Medicine Technology: 38 (4)
Journal of Nuclear Medicine Technology
Vol. 38, Issue 4
December 1, 2010
  • Table of Contents
  • About the Cover
  • Index by author
Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine Technology.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging
(Your Name) has sent you a message from Journal of Nuclear Medicine Technology
(Your Name) thought you would like to see the Journal of Nuclear Medicine Technology web site.
Citation Tools
Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging
Hanna Piwowarska-Bilska, Leszek J. Hahn, Bozena Birkenfeld, Katarzyna Cichon-Bankowska, Maria H. Listewnik, Piotr Zorga
Journal of Nuclear Medicine Technology Dec 2010, 38 (4) 181-185; DOI: 10.2967/jnmt.109.073486

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging
Hanna Piwowarska-Bilska, Leszek J. Hahn, Bozena Birkenfeld, Katarzyna Cichon-Bankowska, Maria H. Listewnik, Piotr Zorga
Journal of Nuclear Medicine Technology Dec 2010, 38 (4) 181-185; DOI: 10.2967/jnmt.109.073486
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
    • Abstract
    • MATERIALS AND METHODS
    • RESULTS
    • DISCUSSION
    • CONCLUSION
    • REFERENCES
  • Figures & Data
  • Info & Metrics
  • PDF

Related Articles

  • No related articles found.
  • PubMed
  • Google Scholar

Cited By...

  • No citing articles found.
  • Google Scholar

More in this TOC Section

  • Early 10-Minute Postinjection [18F]F-FAPI-42 uEXPLORER Total-Body PET/CT Scanning Protocol for Staging Lung Cancer Using HYPER Iterative Reconstruction
  • Single- Versus Dual-Time-Point Imaging for Transthyretin Cardiac Amyloid Using 99mTc-Pyrophosphate
  • Does Arthrography Improve Accuracy of SPECT/CT for Diagnosis of Aseptic Loosening in Patients with Painful Knee Arthroplasty: A Systematic Review and Metaanalysis
Show more Imaging

Similar Articles

SNMMI

© 2025 SNMMI

Powered by HighWire