Journal of Nuclear Medicine Technology
  • SNMMI
    • JNMT
    • JNM
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • My Cart
Journal of Nuclear Medicine Technology

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • Continuing Education
    • JNMT Podcast
    • SNMMI Annual Meeting Abstracts
  • Subscriptions
    • Subscribers
    • Rates
    • Journal Claims
    • Institutional and Non-member
  • Authors
    • Submit to JNMT
    • Information for Authors
    • Assignment of Copyright
    • AQARA Requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
    • Corporate & Special Sales
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • Watch or Listen to JNMT Podcast
  • Visit SNMMI on Facebook
  • Join SNMMI on LinkedIn
  • Follow SNMMI on Twitter
  • Subscribe to JNMT RSS feeds
Research ArticleBrief Communication

Optimization of 89Zr PET Imaging for Improved Multisite Quantification and Lesion Detection Using an Anthropomorphic Phantom

Paul E. Christian, Simon-Peter Williams, Lance Burrell, Paulo Castaneda, Justin Albiani, Nicholas Sandella, Andrei Iagaru, John M. Hoffman, Alex de Crespigny and Sandra Sanabria Bohorquez
Journal of Nuclear Medicine Technology March 2020, 48 (1) 54-57; DOI: https://doi.org/10.2967/jnmt.119.230474
Paul E. Christian
1Genentech, Inc., South San Francisco, California
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Simon-Peter Williams
1Genentech, Inc., South San Francisco, California
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Lance Burrell
2Center for Quantitative Cancer Imaging, Huntsman Cancer Institute, University of Utah School of Medicine, Salt Lake City, Utah
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Paulo Castaneda
3Division of Nuclear Medicine and Molecular Imaging, Department of Radiology, Stanford University, Palo Alto, California; and
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Justin Albiani
4Invicro, Boston, Massachusetts
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Nicholas Sandella
4Invicro, Boston, Massachusetts
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Andrei Iagaru
3Division of Nuclear Medicine and Molecular Imaging, Department of Radiology, Stanford University, Palo Alto, California; and
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
John M. Hoffman
2Center for Quantitative Cancer Imaging, Huntsman Cancer Institute, University of Utah School of Medicine, Salt Lake City, Utah
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Alex de Crespigny
1Genentech, Inc., South San Francisco, California
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Sandra Sanabria Bohorquez
1Genentech, Inc., South San Francisco, California
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

Abstract

Our objective was to harmonize multicenter 89Zr PET imaging for oncology trials and to evaluate lesion detection. Methods: Seven PET scanners were evaluated using a custom chest oncology phantom with 9 spheric lesions 7–20 mm in diameter. A 4:1 signal-to-background ratio simulated a patient dose of 92.5 MBq. Various image reconstructions were evaluated. Images were assessed for lesion detection, and recovery coefficients and background signal variance were measured. Results: Two scanners failed to provide acceptable images and data. Optimal reconstruction algorithms enabling adequate lesion detection and reliable quantification across the other 5 scanners were determined without compromising the data quality. On average, 95% of the 10-mm lesions were detected, and the 7-mm lesion was visualized by only 1 scanner. Background variance was 8.6%–16%. Conclusion: We established multicenter harmonization procedures for 89Zr PET imaging in oncology, optimizing small-lesion (≥10 mm) detectability and accurate quantification.

  • scanner harmonization
  • scanner validation
  • 89Zr PET imaging

PET imaging of radiolabeled monoclonal antibodies can support the development of cancer immunotherapeutics by providing whole-body biodistribution and kinetic information on normal tissue and cancer lesions. PET scanner validation studies by organizations such as the American College of Radiology Imaging Network and EANM Research Ltd. use only technical phantoms with relatively high signal-to-background ratios (SBRs) and with high counts, as they are generally performed with 18F. Thus, they cannot be directly applied for scanner qualification for 89Zr imaging. 89Zr (78.4-h half-life) enables imaging of monoclonal antibodies over the biologically relevant period of several days (1–3). Multicenter imaging settings can accelerate implementation of clinical trials provided that scanners of different makes, models, and technologies are suitably qualified and that standardized imaging procedures are established to ensure that imaging data are qualitatively and quantitatively comparable across sites. Harmonization procedures for 18F-labeled tracers are relatively well established (4,5). However, 89Zr remains challenging because of its much lower count statistics (22.3% positron abundance), its nonprompt 909-keV γ-rays, and the low patient doses typically administered.

Furthermore, the high repeatability coefficients, varying up to 42% in lesions, estimated from clinical 89Zr PET studies (6) further substantiate the need for multisite imaging harmonization. Soderlund et al. (7) demonstrated good-quality images with longer-lived PET radionuclides than 18F but did not optimize image acquisition for doses likely to be used in clinical studies (37–92.5 MBq). Makris et al. (8) and Kaalep et al. (9) addressed the need for 89Zr imaging multicenter harmonization using National Electrical Manufacturers Association body phantoms but with a relatively high 10:1 SBR. These 3 studies, however, failed to evaluate image suitability for interpretation and lesion detection.

Here, we evaluated scanner and reconstruction performance using a chest oncology simulator phantom with 89Zr at clinically relevant doses and a 4:1 SBR. The optimal image reconstruction for each scanner was selected on the basis of small-lesion detection and quantitative performance.

MATERIALS AND METHODS

89Zr phantom imaging was performed with 7 PET scanners at 3 centers: DST, 600, 690, and 710 (GE Healthcare), and EXACT HR+, Biograph 6, and mCT (Siemens). A custom anthropomorphic oncology phantom representing a human chest with uniform body compartment and lung regions was used (Fig. 1). The phantom contained 9 fillable spheric lesions connected in series via small-bore tubing (1 × 7 mm, 5 × 10 mm, 2 × 15 mm, and 1 × 20 mm) (10,11). The phantom was filled to represent a 63-kg patient injected with 92.5 MBq of 89Zr. The body compartment was filled with approximately 1.46 kBq of 89Zr per milliliter. Some of this activity partially filled the lungs. The lesions were filled to create a 4:1 lesion SBR. The radioactive solutions were freshly prepared from approximately 37 MBq (1 mCi) of 89Zr supplied in 5 mL of 1 M oxalic acid, giving final oxalic acid concentrations of approximately 0.6 mM in the lesions and approximately 0.15 mM in the background. Three bed positions were acquired at 10 min per position as might be done clinically. Attenuation correction was performed using low-energy CT or a transmission scan in the HR+ scanner. Various reconstructions were performed on each scanner, using 2–4 iterations, 14–24 subsets, and 4- to 7-mm gaussian filters, as vendor software would allow. Reconstructed images were evaluated by both visual and quantitative methods. Image analysis was performed using OsiriX MD software (Pixmeo SARL).

FIGURE 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 1.

Anthropomorphic chest oncology simulator with 9 internal spheric lesions. Lesion locations and size are shown at left. Phantom positioned supine in scanner is shown at right.

Visual Image Analysis

All image sets were reviewed, and one image set from each scanner was selected for analysis by an experienced operator. This selected image set had the least noise, and unequivocal visualization of at least four 10-mm and larger lesions, and was deemed suitable for radiologist review and reliable interpretation if used clinically. For a given scanner, this reconstruction set allowing the identification of the greatest number of lesions was selected. Although an 18F phantom image and the CT scan were used to verify lesion locations, spheric volumes of interest were delineated only on visually identified lesions. Figure 2 displays the thick-slice maximum-intensity-projection 18F reference images on a 710 scanner, showing the position of all lesions except the 7-mm lesion, which was visible only on the thin-slice images.

FIGURE 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 2.

18F reference and 89Zr maximum-intensity-projection images. The 7-mm lesion was seen only on 710 thin-slice images (not shown). Circles show areas with focal 89Zr uptake not related to lesions.

Quantitative Image Analysis

Quantitative analysis was performed on the sets selected from each scanner in the visual assessment step. A background volume of interest approximately 25 mm in diameter was assigned in a uniform right shoulder area. The average SUV in the background volume of interest (SUVmean) and the SD were recorded and the percentage variation coefficient was calculated (100 × SD/SUVmean) and used as a measure of noise in the selected background region. In lesion volumes of interest, the SUVmax and the average SUV within a 1 cm3 spheric volume (SUVpeak) were used to create recovery coefficient (RC) curves. RC was defined as the ratio of activity concentration measured in the images to the activity concentration injected in the spheres. The measurements in the 10- and 15-mm lesions in the mediastinal region with a 4:1 SBR were used in the RC curves.

RESULTS

Visual Analysis

CT images were used to verify that lesions were uniformly filled, without bubbles. Some PET images showed minor focal 89Zr uptake on an inside phantom surface (Fig. 2), but its location away from lesions did not interfere with the visual and quantitative analysis. This focal uptake was not present during preliminary tests or in previously unused phantoms. It is unclear if this uptake was potentially mediated by biofilm contamination or could have been avoided with additional stabilizing agents (e.g., ethylenediaminetetraacetic acid).

All reconstructed image sets were visually reviewed, and the best image set for each scanner was selected and used for quantitative analysis. The selected reconstruction set parameters for these 5 scanners are given in Table 1.

View this table:
  • View inline
  • View popup
TABLE 1

Optimal Reconstruction Parameters for 89Zr Oncology Imaging

Table 2 shows the lesions visualized per scanner. The 20-mm lesion was visualized by all scanners, the 15- and 10-mm lesions were detected in 6 scanners and 5 scanners, respectively, and the 7-mm lesion was detected in the 710 scanner only. Visual review revealed that the Biograph 6 and HR+ images had the poorest noise characteristics: none of the 10-mm lesions were visualized, and the larger lesions were blurred, with low SBR. Figure 2 shows the thick-slice 710 18F reference and 89Zr maximum-intensity-projection images from all scanners. All lesions are included in these thick-slice maximum-intensity-projection images.

View this table:
  • View inline
  • View popup
TABLE 2

Lesion Visualization

Quantitative Analysis

The background-noise percentage-variation coefficients were 12.1% for the DST, 15.9% for the 600, 15.8% for the 690, 13.1% for the 710, 20.4% for the HR+, 36.4% for the Biograph 6, and 8.6% for the mCT. The SUVpeak RC curves were smoother than the SUVmax curves; therefore, only the SUVpeak curves were used for scanner harmonization and included in the results. The SUVpeak RC percentage decrease between 20-mm and 10-mm lesions calculated as 100% (SUVpeak 20 mm/SUVpeak 10 mm – 1) is shown in Figure 3. The largest decreases were observed for the Biograph 6 and HR+ scanners, at 99.7% and 120.3%, respectively, whereas the drop for all other scanners was less than 74%.

FIGURE 3.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 3.

Percentage reduction of SUVpeak between 20- and 10-mm lesions.

Correlation of Visual and Quantitative Results

The Biograph 6 and HR+ scanners failed to perform adequately either visually or quantitatively; therefore, they were excluded from further analysis. Considering the low count density of the 89Zr images, the SUVpeak RC curves for the 5 remaining scanners showed reasonable agreement (Fig. 4). These scanners—DST, 600, 690, 710, and mCT—had similar quantitative performance and visual characteristics for noise and small-lesion detection. Lesion visualization performance was 100% for 20-mm lesions, 100% for 15-mm lesions, and 95% for 10-mm lesions. The 7-mm lesion was clearly visualized only with the 710 scanner.

FIGURE 4.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 4.

SUVpeak RC curves for 5 harmonized scanners.

DISCUSSION

Several papers have investigated the limitations and accuracy of low-count PET imaging. Kadrmas and Christian (12) showed that reconstruction methods have a significant impact on lesion detection and, thus, that scanning duration and reconstruction algorithms need to be adapted to the imaging conditions. This was confirmed by Carlier et al. (13), who found that 90Y PET imaging in phantoms and patients resulted in image quality and quantitative measurements that varied with the reconstruction algorithms. Work by Makris et al. (8) and Kaalep et al. (9) addressed the need for 89Zr imaging multicenter harmonization using a uniform cylinder and National Electrical Manufacturers Association NU-2 phantoms. However, the large 10:1 SBR and very low background level used in these 2 studies may not be adequate to address the question of lesion visualization and quantification of less avid lesions likely to be observed during this type of study because of the persistent nature of monoclonal antibodies in most tissues.

We evaluated the 89Zr imaging performance of 7 PET scanners using an anthropomorphic phantom with a low radioactive dose and a 4:1 SBR. The recommended reconstruction methods enable optimal lesion detection and quantification. On the basis of the visual and quantitative analysis, the Biograph 6 and HR+ scanners had a poor performance in this low-count imaging situation. These scanners have hardware that is inherently less sensitive than scanners of more recent technology and are limited by their reconstruction algorithms.

Scanner validation and harmonization studies should be performed with the radionuclide being used in the clinical protocol, because the performance of scatter fraction modeling, half-life, randoms, and other factors may influence the reconstructed image quality and quantification.

CONCLUSION

We established a procedure for grading and harmonizing scanners and reconstruction parameters using a 89Zr-filled chest simulator phantom. Scanner-specific reconstruction parameters enabling comparable optimal performance and small-lesion (≥10 mm) detection were determined in 5 scanners to be used in clinical 89Zr-monoclonal antibody PET imaging studies.

DISCLOSURE

Paul E. Christian is a Genentech, Inc., consultant. Simon-Peter Williams, Alex de Crespigny, and Sandra Sanabria Bohorquez are Genentech, Inc., employees and own stock in Roche. Justin Albiani and Nicholas Sandella are Invicro LLC employees. All work was funded by Genentech Inc. No other potential conflict of interest relevant to this article was reported.

Acknowledgments

We thank Keith Bigham of Medical Designs, Inc., Newtown, CT, for manufacturing the custom phantoms.

Footnotes

  • Published online Oct. 11, 2019.

REFERENCES

  1. 1.↵
    1. Verel I,
    2. Visser GW,
    3. van Dongen GA
    . The promise of immuno-PET in radioimmunotherapy. J Nucl Med. 2005;46(suppl):164S–171S.
    OpenUrlAbstract/FREE Full Text
  2. 2.
    1. Zalutsky MR
    . Potential of immuno-positron emission tomography for tumor imaging and immunotherapy planning. Clin Cancer Res. 2006;12:1958–1960.
    OpenUrlFREE Full Text
  3. 3.↵
    1. Reichert JM
    . Monoclonal antibodies as innovative therapeutics. Curr Pharm Biotechnol. 2008;9:423–430.
    OpenUrlCrossRefPubMed
  4. 4.↵
    1. Westerterp M,
    2. Pruim J,
    3. Oyen W,
    4. et al
    . Quantification of FDG PET studies using standardised uptake values in multi-centre trials: effects of image reconstruction, resolution and ROI definition parameters. Eur J Nucl Med Mol Imaging. 2007;34:392–404.
    OpenUrlCrossRefPubMed
  5. 5.↵
    1. Sunderland JJ,
    2. Christian PE
    . Quantitative PET/CT scanner performance characterization based upon the Society of Nuclear Medicine and Molecular Imaging Clinical Trials Network oncology clinical simulator phantom. J Nucl Med. 2015;56:145–152.
    OpenUrlAbstract/FREE Full Text
  6. 6.↵
    1. Jauw YWS,
    2. Heijtel DF,
    3. Zijlstra JM,
    4. et al
    . Noise-induced variability of immuno-PET with zirconium-89-labeled antibodies: an analysis based on count-reduced clinical images. Mol Imaging Biol. 2018;20:1025–1034.
    OpenUrl
  7. 7.↵
    1. Soderlund AT,
    2. Chaal J,
    3. Tjio G,
    4. et al
    . Beyond 18F-FDG: characterization of PET/CT and PET/MR scanners for a comprehensive set of positron emitters of growing application—18F, 11C, 89Zr, 124I, 68Ga and 90Y. J Nucl Med. 2015;56:1285–1291.
    OpenUrlAbstract/FREE Full Text
  8. 8.↵
    1. Makris NE,
    2. Boellaard R,
    3. Visser EP,
    4. et al
    . Multicenter harmonization of 89Zr PET/CT performance. J Nucl Med. 2014;55:264–267.
    OpenUrlAbstract/FREE Full Text
  9. 9.↵
    1. Kaalep A,
    2. Huisman M,
    3. Sera T,
    4. et al
    . Feasibility of PET/CT system performance harmonisation for quantitative multicentre 89Zr studies. EJNMMI Phys. 2018;5:26.
    OpenUrl
  10. 10.↵
    1. Christian PE
    . Use of a precision fillable clinical simulator phantom for PET/CT scanner validation in multi-center clinical trials: the SNM Clinical Trials Network (CTN) program [abstract]. J Nucl Med. 2012;53(suppl 1):437.
    OpenUrl
  11. 11.↵
    1. Christian PE
    . Longitudinal PET scanner stability: SNMMI Clinical Trials network experience [abstract]. J Nucl Med. 2014;55(suppl 1):2156.
    OpenUrl
  12. 12.↵
    1. Kadrmas DJ,
    2. Christian PE
    . Comparative evaluation of lesion detectability for 6 PET imaging platforms using a highly reproducible whole-body phantom with 22Na lesions and location ROC analysis. J Nucl Med. 2002;43:1545–1554.
    OpenUrlAbstract/FREE Full Text
  13. 13.↵
    1. Carlier T,
    2. Willowson KP,
    3. Fourkal E,
    4. et al
    . 90Y-PET imaging: exploring limitations and accuracy under conditions of low counts and high random fraction. Med Phys. 2015;42:4295.
    OpenUrlCrossRefPubMed
  • Received for publication April 30, 2019.
  • Accepted for publication August 12, 2019.
PreviousNext
Back to top

In this issue

Journal of Nuclear Medicine Technology: 48 (1)
Journal of Nuclear Medicine Technology
Vol. 48, Issue 1
March 1, 2020
  • Table of Contents
  • About the Cover
  • Index by author
Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine Technology.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Optimization of 89Zr PET Imaging for Improved Multisite Quantification and Lesion Detection Using an Anthropomorphic Phantom
(Your Name) has sent you a message from Journal of Nuclear Medicine Technology
(Your Name) thought you would like to see the Journal of Nuclear Medicine Technology web site.
Citation Tools
Optimization of 89Zr PET Imaging for Improved Multisite Quantification and Lesion Detection Using an Anthropomorphic Phantom
Paul E. Christian, Simon-Peter Williams, Lance Burrell, Paulo Castaneda, Justin Albiani, Nicholas Sandella, Andrei Iagaru, John M. Hoffman, Alex de Crespigny, Sandra Sanabria Bohorquez
Journal of Nuclear Medicine Technology Mar 2020, 48 (1) 54-57; DOI: 10.2967/jnmt.119.230474

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Optimization of 89Zr PET Imaging for Improved Multisite Quantification and Lesion Detection Using an Anthropomorphic Phantom
Paul E. Christian, Simon-Peter Williams, Lance Burrell, Paulo Castaneda, Justin Albiani, Nicholas Sandella, Andrei Iagaru, John M. Hoffman, Alex de Crespigny, Sandra Sanabria Bohorquez
Journal of Nuclear Medicine Technology Mar 2020, 48 (1) 54-57; DOI: 10.2967/jnmt.119.230474
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
    • Abstract
    • MATERIALS AND METHODS
    • RESULTS
    • DISCUSSION
    • CONCLUSION
    • DISCLOSURE
    • Acknowledgments
    • Footnotes
    • REFERENCES
  • Figures & Data
  • Info & Metrics
  • PDF

Related Articles

  • No related articles found.
  • PubMed
  • Google Scholar

Cited By...

  • No citing articles found.
  • Google Scholar

More in this TOC Section

  • Patient Motion During Cardiac PET Imaging
  • Logistics of Adopting 177Lu-Vipivotide Tetraxetan Therapy in a Community-Based Hospital Setting
  • Fitness for Purpose of Text-to-Image Generative Artificial Intelligence Image Creation in Medical Imaging
Show more Brief Communication

Similar Articles

Keywords

  • Scanner harmonization
  • Scanner validation
  • 89Zr PET imaging