Abstract
99mTc-labeled dimercaptosuccinic acid (99mTc-DMSA) imaging is a well-established and highly sensitive method for the diagnosis of several renal cortical disorders affecting children and adults. Beginning in 2014, 99mTc-DMSA availability was severely impaired when it was added to the Drug Shortages List of the U.S. Food and Drug Administration and was commercially unavailable thereafter. The agent shortage negatively impacted practitioners’ ability to evaluate renal cortical defects in children and adults and changed renal imaging practice. A survey among pediatric nuclear medicine clinicians confirmed the clinical need for 99mTc-DMSA. Finally, in early 2023 the Food and Drug Administration again approved 99mTc-DMSA in the United States. During the 99mTc-DMSA shortage, established practitioners may not have had the opportunity of using 99mTc-DMSA as they were accustomed in their experience. Also, newer imaging specialists and referring physicians and technologists may not have benefited from having 99mTc-DMSA in their training. Therefore, it is time to bring back 99mTc-DMSA into the armamentarium of imaging methods available to evaluate regional cortical renal function.
For the diagnosis and follow-up of several renal disorders in children and adults, 99mTc-labeled dimercaptosuccinic acid (99mTc-DMSA) renal imaging is a well-established method that is of great value (Table 1). It is safe and simple to perform and has a very high diagnostic sensitivity (>90%) for the detection of renal cortical defects (1,2). The renal cortex is where blood is filtered in the nephrons. The filtering unit of the nephron is the glomerulus; it is attached to the convoluted tubule, which removes waste while returning needed substances to the body. Scintigraphic defects on 99mTc-DMSA imaging indicate loss of function in the renal cortex, most commonly due to pyelonephritis or scarring.
99mTc-DMSA has been widely available for decades. Unfortunately, during the period 2014 until 2023, 99mTc-DMSA had a severely limited availability in the United States and was commercially unavailable. The supply interruption was “due to the inability of [a] proposed supplier to reliably produce one of the product’s key components and the additional complexities related to the manufacturing relocation of the final product…” (3). In contrast 99mTc-DMSA continued to be available in Europe and many other countries around the world.
The shortage of 99mTc-DMSA in the United States significantly limited the use of this unique, highly sensitive imaging method. During the shortage of 99mTc-DMSA, only non–Food and Drug Administration–approved 99mTc-DMSA formulations were available in the United States through a very small patchwork of local suppliers. This was not sufficient to meet clinical demand (1,4).
This shortage created a vicious circle in which many nuclear medicine units were no longer performing 99mTc-DMSA scintigraphy due to the unavailability of the radiopharmaceutical, resulting in a significant reduction of clinical demand. Additional factors contributing to its reduced use include older practice guidelines, lack of education and experience of referring physicians about its value, concern about radiation exposure, and others. Understandably, clinicians faced with questions of renal cortical integrity used other imaging techniques despite their lower sensitivity (e.g., ultrasound) and in some cases greater technical complexity (e.g., MRI).
In 2019, a survey among nuclear medicine clinicians clearly demonstrated an ongoing need for the availability of 99mTc-DMSA (Fig. 1). Respondents reported unfamiliarity with how to access alternative formulations of 99mTc-DMSA that were available through several nuclear pharmacies, including both compounded and manufactured formulations. Some providers had access to 99mTc-DMSA but were reluctant to use less stringently regulated, non–Food and Drug Administration–approved compounded formulations or were not aware that a manufactured option was also available (4).
During the diagnostic assessment of urinary tract infection, some advocate obtaining a voiding cystourethrogram to evaluate for the presence of vesicoureteral reflux, which may be followed by 99mTc-DMSA—that is, a “bottom-up” approach. Others prefer a “top-down” approach—that is, obtaining a 99mTc-DMSA scan initially, followed by a voiding cystourethrogram possibly if there is evidence of renal damage. Whether regional preference is for the top-down approach or the bottom-up approach in the evaluation of urinary tract infection, 99mTc-DMSA imaging is a helpful tool for the diagnosis of renal cortical abnormalities (5–7).
Depending on the patient’s clinical question, 99mTc-DMSA imaging can be a valuable complementary test to other diagnostic procedures, or in some cases it can be useful on its own. When used early and appropriately in the evaluation process, it can shorten the time to diagnosis. A commercial source of 99mTc-DMSA is available again in the United States (NEPHROSCAN; Theragnostics Inc., www.nephroscan.com). As many clinicians seek to reinstate its use, the time is right to review and refresh our knowledge on its pharmacodynamics, biodistribution, technical aspects, and clinical value.
RADIOPHARMACEUTICAL KINETICS AND BIODISTRIBUTION
After intravenous injection, 99mTc-DMSA is principally taken up by the brush borders of the proximal convoluted tubules of the renal cortex. The renal uptake and distribution of tracer closely reflect total renal function and regional renal blood flow (8). In patients with normal kidney function, uptake of the injected activity is approximately 40% at 4 h and 70% at 24 h. The renal uptake gradually increases over time. In patients with normal renal function, a very small proportion of the activity is slowly eliminated into the pelvicalyceal system and the bladder. Otherwise, once the activity reaches the renal cortex, it does not wash out or redistribute over time. By 4 h very little extrarenal activity should be present (Fig. 2) (9).
The administered volume and mass of the agent are rather small compared with those of MRI and CT contrast media, therefore not resulting in a pharmacologic, allergic, or toxic effect. As an example, in a 1-y-old weighing 10 kg, the administered volume is 0.03 mL and the mass is 0.22 mg in practice; the 0.03 mL is diluted with normal saline to facilitate injection. As another example, in a 10-kg, 1-y-old patient having a 99mTc-DMSA scan, the administered volume of 99mTc-DMSA solution is 0.06 mL containing only 0.64 mg of radiopharmaceutical. In comparison, for the same 1-y-old undergoing MRI, 940 mg of gadolinium-diethylenetriaminepentaacetic acid (Magnevist; Bayer) is typically administered in a volume of 2.0 mL. For a contrast-enhanced CT scan, 6,400 mg of ioversol (Optiray 320; Covidien) may be administered in a volume of 20 mL. Thus, the administered volume for 99mTc-DMSA is 20- to 200-fold less and the administered mass is more than a thousandfold less (4,9).
Normal split renal function (i.e., the percentage of the total renal uptake in either the right or the left kidney) assessed by 99mTc-DMSA is in the range of 45%–55%.
The total renal uptake of 99mTc-DMSA (right + left kidney) relates to the patient’s total renal function (9). Therefore, if the patient’s total renal function is diminished, the total renal uptake of 99mTc-DMSA will also be diminished. If one kidney has low function, the contralateral kidney will compensate with increased function and size to maintain normal total renal function (10–12).
In neonates, renal uptake of 99mTc-DMSA may be relatively low due to renal immaturity (9,13,14). Although structurally complete in terms of numbers of nephrons by 36 wk, the newborn kidney is still functionally immature. Renal function undergoes rapid maturation during the first weeks after birth in both term and preterm infants. This low renal uptake is usually accompanied by relatively high body background activity. With maturation, renal uptake gradually reaches a normal level. The kidney-to-background ratio gradually approaches the adult level by approximately 2–3 y of age (15).
ADMINISTERED ACTIVITY
According to the North American Guidelines for Administered Activities in Children and Adolescents, the recommended intravenous administered activity of 99mTc-DMSA is 1.85 MBq/kg (0.05 mCi/kg) with a minimum of 11.1 MBq (0.3 mCi) and a maximum 111 MBq (3.0 mCi) (16–18). Ultimately, the amount of administered activity may vary depending on the nuclear medicine physician’s and technologist’s experience and preferences, the patient’s condition, and the type of instrumentation used.
PATIENT PREPARATION
Patients are usually imaged in the normal state of hydration. There are no restrictions for foods, fluids, or physical activity, except when sedation or general anesthesia is planned. The patient and caregivers should be informed about the intravenous injection, the time interval between injection and imaging time, as well as the need for the patient to remain still for the duration of the imaging. They should be assured not to expect any reactions from the injection. Outpatients need to be informed about the need for the patient to meet discharge criteria after sedation or general anesthesia and the time commitment needed.
With the increased use of SPECT and PET, the use of sedation has increased. The use of sedation and general anesthesia may be less frequent with personnel trained and experienced in the handling and imaging of children. In contrast, some institutions use sedation and general anesthesia protocols routinely in certain age groups. Sedation or general anesthesia should be planned ahead of the patient’s visit to nuclear medicine, including the assessment of the appropriateness for the particular patient. It is important to assess the appropriateness of the patient for sedation or general anesthesia before the study. Proper advance instructions about eating, diet, and any other preparation should be communicated clearly to the patient and family. Importantly, if the patient needs more than one imaging procedure in the same day, sedation scheduling should be appropriately coordinated.
TRACER INJECTION
Best results are obtained when experienced, well-trained nurses, technologists, or physicians are the personnel assigned for the intravenous injection. Ensure that appropriate injection technique and immobilization are used. The usual volume of tracer solution is 0.1–0.5 mL followed by a gentle saline flush (1–3 mL) to clear the tracer from the line (19).
IMAGING AND POSTPROCESSING
Typically, imaging is obtained at 2–4 h after tracer administration, and in some cases imaging at 24 h can be useful if tracer is seen in the pelvicalyceal system in 2- to 4-h imaging. The patient should void before imaging and be able to remain still for image acquisition for approximately 20 min. If that is not possible, immobilization, sedation, or, exceptionally, general anesthesia may be considered.
The patient is positioned on the imaging table supine for planar imaging and SPECT. For pinhole imaging the patient should be in the prone position. Details on image acquisition, processing, and display can be found in the Society of Nuclear Medicine and Molecular Imaging (SNMMI) procedure standard/European Association of Nuclear Medicine practice guideline (18).
Even though careful immobilization can be effective, patient motion can affect the quality of the images. Appropriate motion correction techniques are useful and can recover the information, thus avoiding having to repeat the study acquisition (20). Respiratory motion of the kidneys can affect image quality, and there are methods that can reduce the effect of respiratory motion, rendering improved renal images (21).
In the rare instance that the 2- to 4 h images show tracer in the pelvicalyceal system, the patient may be asked to return the next day for additional images (Fig. 3). This instance may occur in the presence of a very dilated collecting system. After the scan, the patient may resume normal activity.
The choice of imaging protocol varies depending on the physician/technologist experience and preferences as well as the available equipment and may involve both planar imaging (with either parallel-hole or pinhole collimator) or SPECT (Fig. 4). Keeping this in mind, generic descriptions of methodology are provided. The methods discussed are based on the Anger-type scintillation cameras and generally available image processing and displays. Newer and developing instrumentation and software methodology can produce clinically acceptable images at a lower radiation exposure and sometimes at a shorter imaging time (22,23).
Planar Imaging (Parallel-Hole Collimator)
Planar imaging may reveal several relevant anatomic findings including the number, size, and position of the kidneys; large cortical defects; differential/split renal function; infection; congenital abnormalities (duplex kidney upper-/lower-pole function, ectopic kidney, cross-fused renal ectopia, horseshoe kidney); infarction; and trauma. Posterior as well as left and right posterior oblique views (300,000–500,000 counts per view) should be obtained as needed. After the initial images are obtained, the physician may wish to order additional views (obliques, anterior, laterals).
Pinhole Collimation
The pinhole collimator, if available, can produce images of very high spatial resolution that may be particularly useful in discerning small renal abnormalities in infants. The pinole aperture size affects image resolution. The smaller the aperture, the higher the spatial resolution; however, the smallest aperture is not as sensitive. We recommend a small aperture (3 mm) if available, although 4 mm may also be acceptable. The highest resolution is obtained when the patient-to-collimator distance is as small as possible. Image size, spatial resolution, and sensitivity decrease with increased patient-to-collimator distance. Pinhole imaging takes more time than with a parallel-hole collimator. When using the pinhole collimator, a prone position is usually used in small children/infants. If the camera system allows, the pinhole collimator positioned under the imaging table as closely as possible to the kidneys is also acceptable.
SPECT
99mTc-DMSA SPECT may be superior to planar imaging for regional mapping of functioning renal parenchyma, providing additional sensitivity and specificity in patients suspected of having renal cortical abnormalities (Fig. 5) (24–26). SPECT can also provide more accurate results than planar imaging, particularly in pediatric patients and poorly functioning kidneys (27). SPECT permits simultaneous evaluation of images in the transverse, coronal, or sagittal plane. In addition, evaluating rotating volume-rendered images such as maximum intensity projections permits a superior gestalt view of the functional anatomy of the kidneys. 99mTc-DMSA SPECT acquisitions can take 15–20 min, requiring the patient to remain still during that time. SPECT acquisition parameters will vary depending on the equipment and software used. According to the SNMMI procedure standard/European Association of Nuclear Medicine practice guideline (18), SPECT imaging requires 360° of sampling, typically on a 128 × 128 matrix with a multihead camera. For most cameras, 120 views in total (3° spacing) with 15–20 s per view can be used. Alternative angular sampling/view times can be used with more advanced reconstructions.
Hybrid SPECT/CT of 99mTc-DMSA provides both functional and anatomic definition of the kidneys, broadening the scope of SPECT imaging (28). With appropriate automated software, 99mTc-DMSA SPECT can allow for the estimation of renal volumes, which can be compared with normal renal values as a function of age and body weight. This information can aid physicians in evaluating and monitoring renal growth (29).
RADIATION EXPOSURE
The use of radiopharmaceuticals such as 99mTc-DMSA in patients involves exposure to ionizing radiation; therefore, attention to patient radiation exposure must be considered. The justification and optimization of its use in the clinic are best summarized by the SNMMI’s statement on dose: “The right test with the right dose should be given to the right patient at the right time.” (30). The effective dose of radiation to patients with 99mTc-DMSA is in the range of 1.0 mSv, which is about the same as or perhaps slightly lower than that associated with pelvic CT. Therefore, if the information that can be derived from 99mTc-DMSA scintigraphy is considered of significant clinical benefit to the patient, this likely outweighs the potential (31) risk of the relatively small amount of radiation exposure (Fig. 6) (32).
COMPARISON WITH ULTRASOUND, CT, AND MRI
99mTc-DMSA SPECT provides a unique 3-dimensional image of the functioning renal cortex. Ultrasonography does not involve ionizing radiation, is generally available, is relatively inexpensive, and can provide initial useful information about the morphology and location of the kidneys. It is operator-dependent, is poorly sensitive for detection of acute pyelonephritis and renal scarring, and may not provide information about viable cortical renal cells. Abdominal CT results in an effective dose of around 5.0 mSv (4). CT with intravenous contrast provides detailed anatomic information, but contrast has risks including allergic reaction and further damage to already compromised kidneys. MRI does not include ionizing radiation but requires a longer imaging time that in most children requires sedation or in some cases general anesthesia, which carries real patient risks. This may be of particular concern since most patients undergoing these evaluations are under 7 y of age (33). MRI has high spatial resolution and high soft-tissue contrast and allows simultaneous evaluation of the collecting system and renal parenchyma. MR urography is usually available only in specialized pediatric centers and can provide dynamic and quantitative information about renal function.
CONCLUSION
99mTc-DMSA renal imaging remains a very useful and sensitive imaging modality in the evaluation of several disorders affecting the renal cortex. The agent shortage negatively impacted practitioners’ ability to evaluate renal cortical defects in children and adults and changed renal imaging practice. As a result, clinical practitioners—both imaging specialists and referring physicians—may be less familiar with the technology than they once were. Newer personnel—both physicians and technologists—may not have encountered the valuable use of 99mTc-DMSA in their training. Since the shortage of commercially available 99mTc-DMSA ended in 2023 and the radiopharmaceutical is now routinely available, it is time to bring 99mTc-DMSA renal imaging back into the fold of imaging options that should be considered to ensure early diagnosis and potentially reduce unnecessary imaging.
DISCLOSURE
No potential conflict of interest relevant to this article was reported.
ACKNOWLEDGMENTS
We would like to thank the nuclear medicine technologists at Boston Children’s Hospital not only for their commitment to technical excellence but also, along with nurses and child life specialists, for their dedication to care and compassion shown to these little patients.
Footnotes
CE credit: For CE credit, you can access the test for this article, as well as additional JNMT CE tests, online at https://www.snmmilearningcenter.org. Complete the test online no later than September 2027. Your online test will be scored immediately. You may make 3 attempts to pass the test and must answer 80% of the questions correctly to receive 1.0 CEH (Continuing Education Hour) credit. SNMMI members will have their CEH credit added to their VOICE transcript automatically; nonmembers will be able to print out a CE certificate upon successfully completing the test. The online test is free to SNMMI members; nonmembers must pay $15.00 by credit card when logging onto the website to take the test.
Published online Aug. 13, 2024.