Journal of Nuclear Medicine Technology
  • SNMMI
    • JNMT
    • JNM
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • Log out
  • My Cart
Journal of Nuclear Medicine Technology

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • Continuing Education
    • JNMT Podcast
    • SNMMI Annual Meeting Abstracts
  • Subscriptions
    • Subscribers
    • Rates
    • Journal Claims
    • Institutional and Non-member
  • Authors
    • Submit to JNMT
    • Information for Authors
    • Assignment of Copyright
    • AQARA Requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
    • Corporate & Special Sales
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • Watch or Listen to JNMT Podcast
  • Visit SNMMI on Facebook
  • Join SNMMI on LinkedIn
  • Follow SNMMI on Twitter
  • Subscribe to JNMT RSS feeds
Review ArticleTeaching Case Studies

An Unusual Cause of the Ring Artifact on Transaxial CT Images

Ashish Kumar Jha, Nilendu C. Purandare and Venkatesh Rangarajan
Journal of Nuclear Medicine Technology September 2016, 44 (3) 214-215; DOI: https://doi.org/10.2967/jnmt.115.165175
Ashish Kumar Jha
Department of Nuclear Medicine and Molecular Imaging, Tata Memorial Hospital, Parel, Mumbai, India
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Nilendu C. Purandare
Department of Nuclear Medicine and Molecular Imaging, Tata Memorial Hospital, Parel, Mumbai, India
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Venkatesh Rangarajan
Department of Nuclear Medicine and Molecular Imaging, Tata Memorial Hospital, Parel, Mumbai, India
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

Abstract

Artifacts and image quality are two sides of the same coin. The ring artifact is scanner-based and caused mainly by either a miscalibrated element or a defective element of a detector row. We describe a rare cause of the ring artifact that appeared on a transaxial CT image because of a loose electronic contact. To our knowledge, this particular cause of the ring artifact has not been described in literature.

  • transaxial image
  • artifact
  • computed tomography
  • detector module
  • motherboard

A diagnostic outcome from PET/CT imaging is the product of images that combine quality, reliability, and reproducibility. Hence, quality control testing of PET/CT equipment, both at acceptance and periodically, is mandatory to achieve the best results (1). Artifacts on CT images, though independent of PET image acquisition, can still degrade the quality of PET images during processing and interpretation (2). Non–scanner-based CT artifacts are caused mainly by improper selection of acquisition parameters and by metallic implants, patient motion, and improper patient preparation (2,3). Scanner-based artifacts may appear because of a sudden breakdown of the system.

CASE REPORT

We encountered the ring artifact on a transaxial CT image while performing routine PET/CT on a 64-slice time-of-flight Gemini TF 64 scanner (Philips) (Fig. 1A, arrows). The artifact persisted on the CT scan of the uniformity phantom (Fig. 1B, arrows) and remained even after cleaning of the Mylar (DuPont Teijin Films) window with warm water. The service engineer performed the bad-detector test to rule out the possibility that the artifact was caused by defective detector modules. Modules 21 and 26 did not pass the test. Two possibilities for the defect were suspected, either an electronic issue within the detector module or a loose connection between the module and the motherboard. Because the first cause would have required replacing the detector module, we tried to rule out the second cause—a loose connection. Both modules were reset in the motherboard, and the bad-detector test was repeated. This time the test did not reveal a problem with any of the modules, including 21 and 26. We then performed CT scanning of the uniformity phantom and found that the ring artifact had disappeared (Fig. 1D). We repeated the scan to confirm disappearance of the artifact. Before starting clinical imaging, we performed an air calibration test. We resumed clinical imaging after obtaining a satisfactory result from this test, and the ring artifact was not present on the clinical images (Fig. 1C).

FIGURE 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 1.

Ring artifact (arrows) present on transaxial CT image of thorax (A) and uniformity phantom (B). Disappearance of ring artifact from both thorax image (C) and phantom image (D).

DISCUSSION

The CT artifacts described in literature are categorized as physics-based, patient-based, or scanner-based (3). Scanner calibration and periodic maintenance by performing a series of quality control tests can help minimize scanner-based artifacts. The ring artifact—the most common scanner-based artifact—can be seen on transaxial CT images for various reasons, such as improper CT number calibration, the presence of any object during air calibration, defective detector elements, or a defective detector module (2,3). Sometimes the ring artifact can appear for accidental reasons such as a contrast drop on the detector or CT tube (4). The main cause of the ring artifact is a faulty detector element, but in our case the artifact was caused by a loosening of the contact between the module and the motherboard, obstructing data flow. This is an unusual phenomenon and occurred because of compromised environmental conditions: a fluctuation in temperature, increase in humidity, and high dust load in the scanner room. The preceding week’s temperature recordings for the PET/CT room showed significant temperature fluctuations to as high as 36°C because the air conditioner had been switched off at night. The manufacturer prescribes an ambient room temperature ranging from 19°C to 22°C (±1°C) for operation of the Gemini TF 64 (5).

CONCLUSION

The ring artifact on transaxial CT images may be due to an unusual cause such as a loose electronic contact, which does not require expensive replacement of the detector module or significant downtime. This cause of the artifact can be avoided by maintaining the strict environmental conditions prescribed by the manufacturer.

DISCLOSURE

No potential conflict of interest relevant to this article was reported.

Footnotes

  • Published online Oct. 15, 2015.

REFERENCES

  1. 1.↵
    1. Seeram E
    . Image quality. In: Computed Tomography: Physical Principles, Clinical Applications and Quality Control. 2nd ed. Philadelphia, PA: Saunders; 2001:174–199.
  2. 2.↵
    1. Wilting JE,
    2. Timmer J
    . Artefacts in spiral-CT images and their relation to pitch and subject morphology. Eur Radiol. 1999;9:316–322.
    OpenUrlCrossRefPubMed
  3. 3.↵
    1. Barrett JF,
    2. Keat N
    . Artifacts in CT: recognition and avoidance. Radiographics. 2004;24:1679–1691.
    OpenUrlCrossRefPubMed
  4. 4.↵
    1. Jha AK,
    2. Purandare NC,
    3. Shah S,
    4. Agrawal A,
    5. Puranik AD,
    6. Rangarajan V
    . Identification of a unique cause of ring artifact seen in computed tomography trans-axial images. Indian J Nucl Med. 2013;28:232–233.
    OpenUrlCrossRefPubMed
  5. 5.↵
    Philips Medical Systems, Inc. Gemini TF and Gemini LXL system information. Release 3.5, document 459800013831, revision A, November 2009.
  • Received for publication August 5, 2015.
  • Accepted for publication September 8, 2015.
PreviousNext
Back to top

In this issue

Journal of Nuclear Medicine Technology: 44 (3)
Journal of Nuclear Medicine Technology
Vol. 44, Issue 3
September 1, 2016
  • Table of Contents
  • About the Cover
  • Index by author
Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine Technology.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
An Unusual Cause of the Ring Artifact on Transaxial CT Images
(Your Name) has sent you a message from Journal of Nuclear Medicine Technology
(Your Name) thought you would like to see the Journal of Nuclear Medicine Technology web site.
Citation Tools
An Unusual Cause of the Ring Artifact on Transaxial CT Images
Ashish Kumar Jha, Nilendu C. Purandare, Venkatesh Rangarajan
Journal of Nuclear Medicine Technology Sep 2016, 44 (3) 214-215; DOI: 10.2967/jnmt.115.165175

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
An Unusual Cause of the Ring Artifact on Transaxial CT Images
Ashish Kumar Jha, Nilendu C. Purandare, Venkatesh Rangarajan
Journal of Nuclear Medicine Technology Sep 2016, 44 (3) 214-215; DOI: 10.2967/jnmt.115.165175
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
    • Abstract
    • CASE REPORT
    • DISCUSSION
    • CONCLUSION
    • DISCLOSURE
    • Footnotes
    • REFERENCES
  • Figures & Data
  • Info & Metrics
  • PDF

Related Articles

  • No related articles found.
  • PubMed
  • Google Scholar

Cited By...

  • A Bizarre Ring Artifact on SPECT/CT Images
  • Google Scholar

More in this TOC Section

  • High-Sensitivity Troponin Elevation in a Young Woman with Typical Chest Pain: The Heart of the Matter
  • Pulmonary Adenocarcinoma Revealed by Parathyroid Scintigraphy: An Incidental Case to Remember
  • Prominent Right Ventricular Tracer Uptake: A Harbinger of Multivessel Coronary Artery Disease
Show more Teaching Case Studies

Similar Articles

Keywords

  • Transaxial image
  • artifact
  • Computed tomography
  • Detector module
  • motherboard
SNMMI

© 2025 SNMMI

Powered by HighWire