- TABLE 1
Differences in Acquisition Parameters Between Diagnostic CT and Attenuation-Correction CT for SPECT/CT MPI
Parameter Diagnostic CT Attenuation-correction CT Voltage (kV) 120 120 Tube current (mA) Automated (∼300–400) 1.5–33 Rotation time (s) 0.33 1.5–30 Effective mAs 100–130 (dependent on automated mA) 24–50 Slice thickness (mm) Acquired 0.5–1 5–10 Reconstructed 3 5–10 Pitch 0.75–1 1–2