Repeatability of SUV measurements in serial PET

Med Phys. 2011 May;38(5):2629-38. doi: 10.1118/1.3578604.

Abstract

Purpose: The standardized uptake value (SUV) is a quantitative measure of FDG tumor uptake frequently used as a tool to monitor therapeutic response. This study aims to (i) assess the reproducibility and uncertainty of SUV max and SUV mean, due to purely statistical, i.e., nonbiological, effects and (ii) to establish the minimum uncertainty below which changes in SUV cannot be expected to be an indicator of physiological changes.

Methods: Three sets of measurements were made using a GE Discovery STE PET/CT Scanner in 3D mode: (1) A uniform 68Ge 20 cm diameter cylindrical phantom was imaged. Thirty serial frames were acquired for durations of 3, 6, 10, 15, and 30 min. (2) Esser flangeless phantom (Data Spectrum, approximately 6.1 L) with fillable thin-walled cylinders inserts (diameters: 8, 12, 16, and 25 mm; height: approximately 3.8 mm) was scanned for five consecutive 3 min runs. The cylinders were filled with 18FDG with a 37 kBq/cc concentration, and with a target-to-background ratio (T/BKG) of 3/1. (3) Eight cancer patients with healthy livers were scanned approximately 1.5 h post injection. Three sequential 3 min scans were performed for one bed position covering the liver, with the patient and bed remaining at the same position for the entire length of the scan. Volumes of interest were drawn on all images using the corresponding CT and then transferred to the PET images. For each study (1-3), the average percent change in SUV mean and SUV max were determined for each run pair. Moreover, the repeatability coefficient was calculated for both the SUV mean and SUV max for each pair of runs. Finally, the overall ROI repeatability coefficient was determined for each pair of runs.

Results: For the 68Ge phantom the average percent change in SUV max and SUV mean decrease as a function of increasing acquisition time from 4.7 +/- 3.1 to 1.1 +/- 0.6%, and from 0.14 +/- 0.09 to 0.04 +/- 0.03%, respectively. Similarly, the coefficients of repeatability also decrease between the 3 and 30 min acquisition scans, in the range of 10.9 +/- 3.9% - 2.6 +/- 0.9%, and 0.3 +/- 0.1% - 0.10 +/- 0.04%, for the SUV max and SUV mean, respectively. The overall ROI repeatability decreased from 18.9 +/- 0.2 to 6.0 +/- 0.1% between the 3 and 30 min acquisition scans. For the l8FDG phantom, the average percent change in SUV max and SUV mean decreases with target diameter from 3.6 +/- 2.0 to 1.5 +/- 0.8% and 1.5 +/- 1.3 to 0.26 +/- 0.15%, respectively, for targets from 8-25 mm in diameter and for a region in the background (BKG). The coefficients of repeatability for SUV max and SUV mean also decrease as a function of target diameter from 7.1 +/- 2.5 to 2.4 +/- 0.9 and 4.2 +/- 1.5 to 0.6 +/- 0.2, respectively, for targets from 8 mm to BKG in diameter. Finally, overall ROI repeatability decreased from 12.0 +/- 4.1 to 13.4 +/- 0.5 targets from 8 mm to BKG in diameter. Finally, for the measurements in healthy livers the average percent change in SUVmax and SUV mean were in the range of 0.5 +/- 0.2% - 6.2 +/- 3.9% and 0.4 +/- 0.1 and 1.6 +/- 1%, respectively. The coefficients of repeatability for SUV max and SUV men are in the range of 0.6 +/- 0.7% - 9.5 +/- 12% and 0.6 +/- 0.7% - 2.9 +/- 3.6%, respectively. The overall target repeatability varied between 27.9 +/- 0.5% and 41.1 +/- 1.0%.

Conclusions: The statistical fluctuations of the SUV mean are half as large as those of the SUV max in the absence of biological or physiological effects. In addition, for clinically applicable scan durations (i.e., approximately 3 min) and FDG concentrations, the SUV max and SUV mean have similar amounts of statistical fluctuation for small regions. However, the statistical fluctuations of the SUVmean rapidly decrease with respect tothe SUVmax as the statistical power of the data grows either due to longer scanning times or as the target regions encompass a larger volume.

Publication types

  • Evaluation Study

MeSH terms

  • Algorithms*
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Neoplasms / diagnostic imaging*
  • Phantoms, Imaging
  • Positron-Emission Tomography / instrumentation
  • Positron-Emission Tomography / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Subtraction Technique*