Molecular breast imaging: use of a dual-head dedicated gamma camera to detect small breast tumors

AJR Am J Roentgenol. 2008 Dec;191(6):1805-15. doi: 10.2214/AJR.07.3693.

Abstract

Objective: Molecular breast imaging with a single-head cadmium zinc telluride (CZT) gamma camera has previously been shown to have good sensitivity for the detection of small lesions. To further improve sensitivity, we developed a dual-head molecular breast imaging system using two CZT detectors to simultaneously acquire opposing breast views and reduce lesion-to-detector distance. We determined the incremental gain in sensitivity of molecular breast imaging with dual detectors.

Subjects and methods: Patients with BI-RADS category 4 or 5 lesions < 2 cm that were identified on mammography or sonography and scheduled for biopsy underwent molecular breast imaging as follows: After injection of 740 MBq of technetium-99m ((99m)Tc) sestamibi, 10-minute craniocaudal and mediolateral oblique views of each breast were acquired. Blinded reviews were performed using images from both detectors 1 and 2 and images from detector 1 only (simulating a single-head system). Lesions were scored on a scale of 1-5; 2 or higher was considered positive.

Results: Of the 150 patients in the study, 128 cancers were confirmed in 88 patients. Averaging the results from the three blinded readers, the sensitivity of dual-head molecular breast imaging was 90% (115/128), whereas the sensitivity from review of only single-head molecular breast imaging was 80% (102/128). The sensitivity for the detection of cancers < or = 10 mm in diameter was 82% (50/61) for dual-head molecular breast imaging and 68% (41/61) for single-head molecular breast imaging. On average, 13 additional cancers were seen on dual-head images and the tumor uptake score increased by 1 or more in 60% of the identified tumors.

Conclusion: Gains in sensitivity with the dual-head system molecular breast imaging are partially due to increased confidence in lesion detection. Molecular breast imaging can reliably detect breast lesions < 2 cm and dual-head molecular breast imaging can significantly increase sensitivity for subcentimeter lesions.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Breast Neoplasms / diagnostic imaging*
  • Equipment Design
  • Equipment Failure Analysis
  • Female
  • Gamma Cameras*
  • Humans
  • Image Enhancement / instrumentation*
  • Middle Aged
  • Molecular Probe Techniques / instrumentation*
  • Radionuclide Imaging
  • Radiopharmaceuticals
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Single-Blind Method
  • Technetium Tc 99m Sestamibi*

Substances

  • Radiopharmaceuticals
  • Technetium Tc 99m Sestamibi