Monte carlo simulations of dose from microCT imaging procedures in a realistic mouse phantom

Med Phys. 2006 Jan;33(1):216-24. doi: 10.1118/1.2148333.

Abstract

The purpose of this work was to calculate radiation dose and its organ distribution in a realistic mouse phantom from micro-computed tomography (microCT) imaging protocols. CT dose was calculated using GATE and a voxelized, realistic phantom. The x-ray photon energy spectra used in simulations were precalculated with GATE and validated against previously published data. The number of photons required per simulated experiments was determined by direct exposure measurements. Simulated experiments were performed for three types of beams and two types of mouse beds. Dose-volume histograms and dose percentiles were calculated for each organ. For a typical microCT screening examination with a reconstruction voxel size of 200 microm, the average whole body dose varied from 80 mGy (at 80 kVp) to 160 mGy (at 50 kVp), showing a strong dependence on beam hardness. The average dose to the bone marrow is close to the soft tissue average. However, due to dose nonuniformity and higher radiation sensitivity, 5% of the marrow would receive an effective dose about four times higher than the average. If CT is performed longitudinally, a significant radiation dose can be given. The total absorbed radiation dose is a function of milliamperes-second, beam hardness, and desired image quality (resolution, noise and contrast). To reduce dose, it would be advisable to use the hardest beam possible while maintaining an acceptable contrast in the image.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Animals
  • Body Burden
  • Computer Simulation
  • Mice
  • Models, Biological*
  • Models, Statistical
  • Monte Carlo Method
  • Phantoms, Imaging
  • Radiation Dosage
  • Radiometry / methods*
  • Relative Biological Effectiveness
  • Tomography, X-Ray Computed / methods*
  • Tomography, X-Ray Computed / veterinary*
  • Whole Body Imaging / methods*
  • Whole Body Imaging / veterinary*