Elsevier

Journal of Nuclear Cardiology

Volume 6, Issue 5, September–October 1999, Pages 498-506
Journal of Nuclear Cardiology

Original article
Automatic quantification of left ventricular ejection fraction from gated blood pool SPECT

https://doi.org/10.1016/S1071-3581(99)90022-3Get rights and content

Abstract

Background. Cardiac gated blood pool single photon emission computed tomography (GBPS) better separates cardiac chambers compared with planar radionuclide ventriculography (PRNV). We have developed a completely automatic algorithm to measure quantitatively the left ventricular ejection fraction (LVEF) from gated technetium 99m-red blood cells (RBC) GBPS short-axis 3-dimensional image volumes.

Methods and Results. The algorithm determines an ellipsoidal coordinate system for the left ventricle and then computes a static estimate of the endocardial surface by use of counts and count gradients. A dynamic surface representing the endocardium is computed for each interval of the cardiac cycle by use of additional information from the temporal Fourier transform of the image data sets. The algorithm then calculates the left ventricular volume for each interval and computes LVEF from the end-diastolic and end-systolic volumes. The algorithm was developed in a pilot group (N = 45) and validated in a second group (N = 89) of patients who underwent PRNV and 8-interval GBPS. Technically inadequate studies (N = 38) were rejected before grouping and processing. Automatic identification and contouring of the left ventricle was successful in 121172 patients (70%) globally and in 7689 patients (85%) in the validation group. Correlation between LVEFs measured from GBPS and PRNV was high (y = 2.00 + 1.01x, r = 0.89), with GBPS LVEF significantly higher than PRNV LVEF (average difference = 2.8%, P < .004).

Conclusions. Our automatic algorithm agrees with conventional radionuclide measurements of LVEF and provides the basis for 3-dimensional analysis of wall motion.

References (44)

  • Risk stratification and survival after myocardial infarction

    N Engl J Med

    (1983)
  • KL Lee et al.

    Prognostic value of radionuclide angiography in medically treated patients with coronary artery disease: a comparison of clinical and catheterization variables

    Circulation

    (1990)
  • DS Berman et al.

    Clinical assessment of left ventricular regional contraction patterns and ejection fraction by high-resolution gated scintigraphy

    J Nucl Med

    (1974)
  • JS Borer et al.

    Real-time radionuclide cineangiography in the noninvasive evaluation of global and regional left ventricular function at rest and during exercise in patients with coronary-artery disease

    N Engl J Med

    (1977)
  • J Maddahi et al.

    A new technique for assessing right ventricular ejection fraction using rapid multiple-gated equilibrium cardiac blood pool scintigraphy: description, validation and findings in chronic coronary artery disease

    Circulation

    (1979)
  • ML Moore et al.

    ECG-gated emission computed tomography of the cardiac blood pool

    Radiology

    (1980)
  • W Chang et al.

    Seven-pinhole multigated tomography and its application to blood-pool imaging: technical parameters

    J Nucl Med

    (1980)
  • JB Gill et al.

    Multigated blood pool tomography: new method for the assessment of left ventricular function

    J Nucl Med

    (1986)
  • M Bartlett et al.

    SPECT gated blood pool imaging—potentials and problems

  • BM Massie et al.

    Radionuclide measurement of left ventricular volume: comparison of geometric and counts-based methods

    Circulation

    (1982)
  • R Slutsky et al.

    Left ventricular volumes by gated equilibrium radionuclide angiography: a new method

    Circulation

    (1979)
  • RF Uren et al.

    Geometric determination of left ventricular volume from gated blood pool studies using a slant-hole collimator

    Radiology

    (1983)
  • Cited by (98)

    • Evaluation of the right ventricle: Comparison of gated blood-pool single photon electron computed tomography and echocardiography with cardiac magnetic resonance

      2014, International Journal of Cardiology
      Citation Excerpt :

      A triple-head gamma camera (Prism 3000S, Picker Corporation, Cleveland, Ohio) was used to collect images in 63 projections over a 360° ellipsoidal arc. QBS ventricular segmentation was performed by applying a ventricular mask that excludes activity from the atria and extra-cardiac structures [29]. Ventricular surfaces were spatially and temporally used to compute RVEDV, RVESV, and RVEF.

    View all citing articles on Scopus