Skip to main content

Advertisement

Log in

A novel method to localize antibody-targeted cancer deposits intraoperatively using handheld PET beta and gamma probes

  • Published:
Surgical Endoscopy Aims and scope Submit manuscript

Abstract

Background

Assessing cancer margins, lymph nodes, and small cancer deposits intraoperatively can be challenging. A new device has become available that allows the detection of positron emission tomography (PET) radiotracers through both high-energy gamma and short-range beta emissions. These PET probes are handheld, allowing for real-time evaluation of cancer using a tool that provides surgeons with better intraoperative assessment of tumor sites.

Methods

Within the context of two institutional review board (IRB)-approved protocols investigating new applications of antibody-labeled PET scanning, 124I-labeled humanized monoclonal antibodies specific for colorectal cancer (huA33) and renal tumors (cG250) were constructed. Patients underwent preoperative PET scans, approximately seven days post-tracer infusion, when tumor-to-nontumor ratios were high. Suspected tumor deposits were evaluated intraoperatively with handheld beta and gamma PET probes.

Results

Handheld PET probes detected emissions from all tumors. Count rates from the gamma probe on tumor ranged from 48 to 306 cps, and for the beta probe ranged from 18 to 190 cps. Gamma and beta emissions exhibited a strong positive correlation. The ratio of gamma and beta counts was at least twice that of the background counts for all tumors evaluated.

Conclusions

This study is the first to demonstrate the utility of beta probes for the intraoperative detection of radiolabeled antibodies targeting cancer. Importantly, the recorded beta count rates from the beta probe correlate with the count rates from the high-energy gamma probe. Furthermore, the beta probe may offer superior specificity for real-time localization of small tumor deposits, compared to gamma probes. The intraoperative portable PET probe may prove a valuable bridge to combining tumor biology and PET technology to guide surgical therapy.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Alazraki NP, Mishkin FS (1991) Fundamentals of nuclear medicine, 2nd edition. The Society of Nuclear Medicine, Inc., ISBM? 0-932004-29-6, p 85

  2. Daghighian F, Mazziotta JC, Hoffman EJ, Siegel S, Shenderov P, Eshaghian B, Phelps ME (1994) Intraoperative beta probe: A device for detecting tissue labeled with positron- or electron-emitting isotopes during surgery. Med Phys 21:153–157

    Article  CAS  PubMed  Google Scholar 

  3. Essner R, Daghighian F, Giuliano A (2002) Advances in FDG PET probes in surgical oncology. Cancer J8:100–108

    Google Scholar 

  4. Gulec SA, Dahgighian F, Essner R (2006) PET Probe: Evaluation of technical performance and clinical utility of a handheld high-energy gamma probe in oncologic surgery. Ann Surg Oncol 1–8

  5. Essner R, Hsueh E, Haigh P, Glass E, Huynh Y, Daghighian F (2001) Application of an 18F fluorodeoxyglucose-sensitive probe for the intraoperative detection of malignancy. J Surg Res 96:120–126

    Article  CAS  PubMed  Google Scholar 

  6. Franc B, Mari C, Johnson D, Leong S (2005) The role of a positron and high-energy gamma photon probe in intraoperative localization of recurrent melanoma. Clin Nucl Med 30:787–791

    Article  PubMed  Google Scholar 

  7. Divgi CR, Pandit-Taskar N, Jungbluth AA, Reuter VE, Gonen M, Ruan S, Pierre C, Nagel A, Pryma DA, Humm J, Larson SM, Old LJ, Russo P (2007) Preoperative characterisation of clear-cell carcinoma using iodine-124-labelled antibody chimeric G250 (124I-cG250) and PET in patients with renal masses: a phase I trial. Lancet Oncol 8:304–310

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgments

Supported in part by a grant from the Society of American Gastrointestinal Endoscopic Surgeons (V.S.), and in part by the Ludwig Center for Cancer Immunotherapy and 5 PO1 Ca33049.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Vivian E. Strong.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Strong, V.E., Humm, J., Russo, P. et al. A novel method to localize antibody-targeted cancer deposits intraoperatively using handheld PET beta and gamma probes. Surg Endosc 22, 386–391 (2008). https://doi.org/10.1007/s00464-007-9611-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00464-007-9611-3

Keywords

Navigation