Preamble

The 2015 update of the European procedural guidelines for radionuclide myocardial perfusion imaging (MPI) with single-photon emission computerized tomography (SPECT) was developed on the initiative of the Cardiovascular Committee of the European Association of Nuclear Medicine (EANM) [1]. The guidelines summarize the views of the Cardiovascular Committee of the EANM and reflect recommendations for which the EANM cannot be held responsible. The recommendations should be taken into the context of good practice of nuclear medicine, and they do not substitute for national and international legal or regulatory provisions. The guidelines were brought to the attention of all other EANM Committees and to the National Societies of Nuclear Medicine and have been approved by the EANM.

The present guidelines are based on the guidelines from 2005 [2]. The 2015 update includes all aspects of SPECT imaging from gating to hybrid imaging, but does not include myocardial perfusion evaluated by positron emission tomography (PET), which will be updated in a separate guideline. The 2015 guideline also updates relevant sections of the 2008 guidelines on radionuclide imaging of cardiac function [3] and the 2011 joint position statement on hybrid cardiac imaging [4].

The 2015 EANM guidelines on myocardial perfusion imaging (MPI) include the following sections:

  1. 1.

    Patient information and preparation

  2. 2.

    Radiopharmaceuticals and CT contrast agents

  3. 3.

    Injected activities, dosimetry, and radiation exposure

  4. 4.

    Stress tests

  5. 5.

    Instrumentation

  6. 6.

    Imaging protocols

  7. 7.

    Image acquisition

  8. 8.

    Quality control of instrumentation and image acquisition

  9. 9.

    Reconstruction methods

  10. 10.

    Attenuation and scatter correction

  11. 11.

    Data analysis of regional perfusion imaging

  12. 12.

    Data analysis of left ventricular function

  13. 13.

    Data analysis of hybrid imaging

  14. 14.

    Reports and image display.

All sections have been updated and a lot of details were revisited. The present paper highlights the major changes presented in the updated 2015 guidelines and focuses on the selective coronary vasodilator regadenoson, recent developments on calculated radiation exposure [i.e., new International Commission on Radiological Protection (ICRP) models, dedicated cardiac SPECT cameras and hybrid systems], recent developments in instrumentation (i.e., dedicated cardiac SPECT and cardiac hybrid imaging systems) and coronary CT-contrast agents. All 14 sections are described in more detail in the fully revised version available at the EANM home page [1].

The EANM guidelines are intended to present information specifically adapted to European practice, based on evidence from original scientific studies or on previously published guidelines [i.e., for the 2015 MPI guidelines as well as for the extract presented below: on national or European guidelines for MPI, on the European Society of Cardiology (ESC), and on the American College of Cardiology (ACC) / American Heart Association (AHA) / American Society of Nuclear Cardiology (ASNC) guidelines]. In case of a lack of published evidence, opinions are based on expert consensus and are indicated as such. Where more than one solution seems to be practised, and none has been shown to be superior to the others, the committee has tried to specifically express this state of knowledge.

The selective coronary vasodilator regadenoson

Mechanism of action

Vasodilators induce myocardial hyperaemia mediated by adenosine receptors independent of myocardial oxygen demand. Only A2A receptors induce coronary vasodilation, and hereby a fourfold to fivefold increase in myocardial blood flow in healthy coronary vessels. Besides, the A2A receptor adenosine also stimulates A1, A2B, and A3 adenosine receptors, which provoke the adverse effects [5]. Regadenoson is a selective stimulator of the A2A receptor with minimal or no stimulation of the other adenosine receptor subtypes [5]. Dipyridamole increases the tissue levels of adenosine by preventing the intracellular reuptake and deamination of adenosine [5]. Vasodilatation results in a modest increase in heart rate and most often a modest decrease in both systolic and diastolic blood pressures.

For regadenoson, as with the other vasodilators, caffeine-containing beverages (coffee, tea, cola, etc.), foods (chocolate, etc.), some medicaments (e.g., pain relievers, stimulants and weight control drugs) and methylxanthine-containing medications that antagonise vasodilator action must be discontinued at least 12 h before vasodilator stress and at least five half-lives before for long-acting methylxanthines. Dipyridamole or medications containing dipyridamole should be interrupted for at least 24 h. Pentoxifylline and clopidogrel need not be discontinued. Ticagrelor, a direct-acting P2Y12-adenosine diphosphate receptor blocker, has been shown to significantly raise adenosine plasma levels so that more frequent and more severe adverse effects in adenosine and dipyridamole stress testing are likely [6, 7]. Interactions with regadenoson have not yet been studied. So far there are still no recommendations for dose modification of vasodilators in stress testing for patients receiving ticagrelor.

Indications

The indications are the same as for exercise MPI, but refer to patients who are not able to or who are expected to be unable to achieve ≥ 85 % of maximal age-predicted heart rate during exercise. Vasodilators (without exercise) should be preferred to exercise in cases of left bundle branch block or ventricular paced rhythms (Fig. 1). Considering diagnostic performance of MPI, there is no significant difference among the stress agents and modalities [810].

Fig. 1
figure 1

Selection of stress test modality. Except for patients with left bundle branch block (LBBB) or ventricular paced rhythm, consider combining pharmacological vasodilatory stress with low-level exercise according to the ability of the patient to exercise. In case of pharmacological stress with dobutamine but without adequate heart rate response, consider to add atropine

Combination with low-level exercise

Low-level exercise can be performed routinely in conjunction with vasodilator tests. Low-level exercise significantly reduces vasodilator-induced side effects (flushing, dizziness, nausea, headache, hypotension) and improves image quality due to lower bowel activity and higher target-background ratio. Accordingly, if possible, low-level exercise is recommended in combination with vasodilator stress testing [11, 12].

Regadenoson dose

Regadenoson is given as a slow bolus over 10 s followed by a 10 s flush of 5–10 ml NaCl 0.9 %, and 10–20 s later the radiopharmaceutical is injected (Fig. 2). Regadenoson is administered independent of patient weight in a dose of 0.4 mg in 5 ml.

Fig. 2
figure 2

Timeline of regadenoson plus low-level exercise testing. BPM beats per minute

Regadenoson safety profile

Compared to adenosine, regadenoson is associated with a lower incidence of chest pain, flushing, and throat, neck, or jaw pain, a higher incidence of headache and gastrointestinal discomfort and a lower combined symptom score in nearly all subgroups [13]. The safety of regadenoson was also studied in two randomized, double-blind, placebo-controlled crossover trials in patients with mild to moderate asthma and moderate to severe chronic obstructive pulmonary disease (COPD), respectively. Although dyspnoea was reported in the regadenoson groups compared to none in the placebo groups, there was no relation with a decline in forced expiratory volume in 1 s (FEV1) [14, 15]. Thus, regadenoson can also be used in patients with mild asthma and COPD. In summary, despite their selectivity to the A2A receptor, side effects related to activation of the other adenosine receptors continue to occur, albeit at a lower frequency and with less severity and duration compared to the selective adenosine agonists.

A less frequent but important side effect of regadenoson is a low but significant increase in incidence of seizures. The exact incidence is unknown and the pathophysiological mechanism behind this increased incidence is not yet clear. Regadenoson may lower seizure threshold, and aminophylline should not be used in cases of seizures associated with regadenoson. These seizures may be of new onset, or may be recurrences. In addition, some seizures are prolonged and may require urgent anticonvulsive management. It is recommended that during the initial triaging of patients, they should be asked about a history of seizures [1618].

According to the available limited data, the absolute risk of a transient ischaemic attack (TIA) and cerebrovascular accident (CVA) associated with regadenoson administration appears to be small and may not be different from other stress agents [19]. However, based on recent advice from the European Medicines Agency, a Direct Healthcare Professional Communication updated the regadenoson product information [20]:

  • As aminophylline increases the risk of prolongation of a regadenoson-induced seizure, it should therefore not be administered solely for the purpose of terminating a seizure induced by regadenoson

  • Delay regadenoson administration in patients with elevated blood pressure until the latter is well controlled

  • There is a rare but undesirable risk of a TIA and CVA associated with regadenoson administration.

Radiation exposure: new ICRP models

The ICRP has adopted new phantoms for absorbed dose calculations, new weighting factors, and new bio-kinetic models [2123]. Using these new models, the effective doses per administered unit of activity to adults are [24]:

  • 99mTc-tetrofosmin (stress and rest): 0.0058 and 0.0063 mSv/MBq, respectively

  • 99mTc-sestamibi (stress and rest): 0.0066 and 0.0070 mSv/MBq, respectively

  • 201Tl-chloride (redistribution): 0.102 mSv/MBq.

These values are approximately 20 % lower than the values calculated using the previous methods.

Activity amounts to inject

Camera systems based on new technology (e.g., cadmium-zinc-telluride (CZT)-cameras) have improved count sensitivity. This improved sensitivity can be used to either reduce the amount of activity accordingly, or decrease image acquisition duration [2529]. The activities to be administered for paediatric patients should be modified according to the recommendations of the EANM [30].

Hybrid systems

If the MPI is combined with CT for attenuation, calcium scoring (CACS) or coronary CT angiography (CCTA), an additional dose is given to the patient. For an attenuation correction CT, an additional 0.5–1.0 mSv is given. Absorbed doses from CACS and CCTA depend on the system and imaging protocol used, and for CACS can be estimated to be < 1 mSv. The absorbed dose for CCTA can be estimated to be between 2–5 mSv using commonly available single-source 64-slice CT scanners with a prospectively ECG-triggered, step-and-shoot acquisition protocol [31, 32]. The latest generation dual-source or 256-slice and 320-slice single source CT scanners enable even absorbed doses < 1 mSv [33, 34].

Instrumentation: dedicated cardiac systems

Since the publication of the first version of this guideline, some significant advances in technology for cardiac imaging have been made [3537]. In new-generation, dedicated cardiac, ultrafast-acquisition scanners, multiple detectors surrounding the patient simultaneously image the heart. Variations in these newly designed, dedicated cardiac scanners comprise the number and type of scanning or stationary detectors and the use of NaI, CsI, or CZT solid-state detectors. They all have in common the potential for a fivefold to tenfold increase in count sensitivity at no loss of resolution, resulting in the potential for acquiring a scan in 2 min or less if the patient is injected with a standard activity. Some of this gain in sensitivity can be traded for a linear reduction in the injected activity to reduce the patients’ exposure to radiation. With an ultrafast camera with a tenfold increase in sensitivity using conventional radiopharmaceutical activity, the dose could be reduced by half and a fivefold increase in sensitivity would still be maintained [36].

Dedicated systems have emerged based on traditional scintillation detector technology, equipped with smaller crystals and thus having more compact design, as well as truly innovative cameras with a completely novel detector technology. Pixelated, solid-state detectors made of CdZnTe (an alloy of cadmium telluride and zinc telluride, CZT) offer better energy resolution and a virtually linear count rate response. As CZT replaces both the scintillation crystal and the attached photomultiplier tubes, the detectors are very compact. This has thus far been exploited in two different, commercially available cameras. The D-SPECT® (Spectrum Dynamics) utilises nine small rectangular detectors placed along a 90° arc. Each detector rotates around its own axis. All detectors together register photons from an area comparable to a traditional 180° acquisition. The other CZT system (the Discovery NM530c by GE Healthcare) uses a stationary, multiple-pinhole design with 19 holes, each with its own CZT detector. The collimators are arranged such that the area of the chest including the heart can be imaged. Clinical evaluations of both systems have demonstrated performance similar to that of traditional systems, but with shorter imaging times or lower administered activities [36]. In a recent study, the D-SPECT® system was compared to a conventional SPECT camera, showing improved image quality, comparable incidence of extracardiac activity, and achieving a reduction in absorbed dose to 1 mSv for a single injection [38].

Another variation from traditional design is the upright patient position offered in a number of dedicated systems. While more comfortable for the patient, the effect of posture on organ position must be considered when interpreting the images [39, 40]. Some cameras systems (e.g., Digirad Cardius) go one step further and employ a rotating patient chair instead of a rotating gantry. Additionally, CsI(Tl) detectors allow for an optional X-ray-based attenuation correction method. Another variation of the upright design (the CardiArc system) is based on three stationary arc-shaped crystals combined with a moving aperture, providing a similar function as a collimator. A problem with many different, dedicated cardiac systems is the scarcity of publications about their clinical validity [36].

An upgraded technology (IQ-SPECT, Siemens Healthcare) uses a conventional multi-purpose SPECT system with a radially oriented, symmetric, cardio-focal collimator that is characterised by a radially increasing focal spot distance. The proprietary reconstruction algorithm matches the collimators’ spatially varying sensitivity profile. Reductions in acquisition time with no loss in image quality are claimed, but the clinical evidence is limited.

Instrumentation: SPECT/CT hybrid systems

The SPECT detectors in most SPECT/CT systems do not differ in any significant way from those of stand-alone SPECT systems. The SPECT sub-systems are typically large field of view (FOV), variable-angle, dual detector system. For hybrid imaging systems, the CT configuration can be a low-resolution CT (non-diagnostic CT) or a multi-detector-row CT with slices ranging from 2 up to 64. Any of these systems can be used for attenuation correction of MPI. For CACS, at least 4-slice CT is required, but ≥ 6-slice recommended. For CCTA, at least a 16-slice scanner is required, but a ≥ 64-slice multidetector-row CT is recommended, with an imaging capability for slice width of 0.4–0.6 mm and temporal resolution of 500 ms or less; ≤ 350 ms is preferred [41, 42].

In general, it has to be considered that CT imaging is much faster than SPECT, where the heart position is averaged over the complete acquisition time (5–20 min), so that mis-registration artefacts can occur. As a consequence, free-breathing and end-expiratory breath-hold protocols during CT scanning are preferred over inspiration breath-hold protocols when the CT scan is performed for attenuation correction only [43]. General guidelines for CT-based transmission imaging for SPECT are listed in Table 1 [44].

Table 1 General guidelines for CT-based transmission imaging for SPECT [44]

For attenuation correction of perfusion SPECT studies, separate CT scans should be performed for the rest and stress MPI studies, even if recent studies demonstrate that the CT transmission scans are interchangeable in specific clinical settings [45]. Other recent studies have demonstrated that the CT scan can be used to approximate the extent of coronary calcification [46]. This approach, however, is less accurate as compared to dedicated calcium scans due to missing correction of the coronary motion and a much lower photon density. Otherwise, the coronary artery calcium score CT scan can be used for attenuation correction, but with the limitation that this scan may not register adequately with SPECT due to different acquisition time points [47, 48]. These are topics of ongoing research and software developments.

For detailed acquisition protocols of the coronary artery calcium scan and CCTA we refer to the joint position statement by the EANM, the ESCR and the European Council of Nuclear Cardiology [4].

Coronary CT contrast agents

The use of contrast media for cardiac imaging is increasing as hybrid cardiac SPECT/CT and PET/CT, as well as coronary CT angiography and cardiac MRI, become more widely used [4].

For coronary CT angiography (CCTA), an intravenously injected contrast agent is needed. In general, these contrast agents are iodine-based and due to the relatively high attenuation coefficient of iodine, therefore result in high contrast between organs with and those without contrast. Coronary CT in combination with ECG-gating and a contrast agent permits visualization of the coronary artery lumen and detection of coronary artery stenoses [49, 50]. At the moment, there are four classes of contrast media available for clinical use: high osmolar ionic monomers, low osmolar non-ionic monomers, low osmolar ionic dimers and iso-osmolar non-ionic dimers. The contrast media are provided at various iodine concentrations and have different biochemical properties (viscosity, osmolarity, hydrophilic behaviour, ionic content and pH).

In general, the injection protocol should follow vendor recommendations dependent on the specific contrast agent and CT used. For coronary imaging in general, a contrast agent with a high concentration of iodine is used (300–400 mg/ml) to ensure adequate opacification of the small coronary arteries [51]. In total, approximately 60–90 ml of contrast agent is injected at an injection speed of 4–6 ml/s. Often the bolus is split between a first bolus of pure contrast and then a bolus with a mix of contrast and saline, to reduce streak artefacts from contrast enhancement of the vena cava and right side of the heart. Otherwise, the contrast bolus is followed by a saline flush of 40–70 ml. With the newest generation of CT scanners, a smaller contrast bolus with lower iodine concentration is likely sufficient to obtain good contrast enhancement of the coronary arteries [52].

Two contrast timing techniques are available to start the CCTA acquisition, based on the arrival of contrast in the aorta: the bolus tracking and the bolus timing technique. Bolus tracking involves a series of axial low-dose images at 2 s intervals to track the arrival of the bolus of contrast material in the aorta. The CCTA is initiated when the contrast enhancement of the aorta reaches a certain predefined level, e.g., 100 Hounsfield units (HU). The bolus timing technique involves an extra low-dose scan acquisition of a single slice prior to the CCTA acquisition. Here, a small contrast bolus and saline flush are injected to determine the contrast arrival interval. The time between the start of the contrast injection and the arrival of contrast bolus in the aorta is used as the scan delay for the actual CCTA [4, 49].

Contraindications

Contraindications for iodine contrast can be divided into absolute and relative contraindications [53, 54]. Absolute contraindications include myasthenia gravis, mastocytosis, and post-thyroid carcinoma when follow-up with 131I imaging or 131I therapy is planned within 6 months of the CCTA. Relative contraindications are known contrast allergy, planned thyroid scan, and multiple myeloma.

The most frequent problem is related to renal failure: The clinical benefit of using estimated glomerular filtration rate (eGFR) or calculated creatinine clearance in assessing pre-procedural, contrast induced nephrotoxicity (CIN) risk in patients with stable renal function is uncertain, because much of the knowledge comes from studies that used only serum creatinine measurements. The threshold values at which different clinical actions should be taken (e.g., active IV hydration, avoidance of contrast medium administration) are neither proven nor generally agreed upon for either serum creatinine measurement or calculated creatinine clearance. In addition, the accuracy of these formulae has only been validated in the patient population for whom they were developed. The following is a suggested list of risk factors that may warrant pre-administration serum creatinine screening in patients who are scheduled to receive intravascular iodinated contrast medium. This list should not be considered definitive [55]:

  • Age > 60

  • History of renal disease, including dialysis, kidney transplant, single kidney, renal cancer, renal surgery

  • History of hypertension requiring medical therapy

  • Diabetes.

Metformin does not confer an increased risk of CIN. However, metformin can very rarely lead to lactic acidosis in patients with renal failure. In case of reduced renal function (i.e., estimated glomerular filtration rate < 45 ml/min/1.73 m2, or < 60 ml/min/1.73 m2 in the presence of diabetes or ≥ 2 risk factors for contrast nephropathy), alternative imaging strategies should be considered. However, when no alternatives are available, adequate hydration is the major preventive action against CIN.

If intravenous contrast agent is going to be administered, metformin should be discontinued at the time of the procedure and withheld for 48 h after the procedure. If the risk of nephrotoxicity is high, metformin can be reinstituted only after renal function has been re-evaluated and found to be normal. If the risk of nephrotoxicity is low, metformin can be re-instituted without the need for renal function assessment. An alternative glucose-controlling drug should be considered during this time [56].

Adverse events

Strategies to reduce the risk in non-acute, contrast medium induced, adverse reactions include the prophylactic use of oral anti-histamines and corticosteroid tablets. A classification system, stratifying adverse events due to iodinated contrast media by severity and type, is presented below [55, 57, 58]:

  • Minor Signs and symptoms are self-limited without evidence of progression. Mild reactions include: pruritus, nausea and mild vomiting, diaphoresis

  • Moderate: Signs and symptoms are more pronounced and commonly require medical management. Some of these reactions have the potential to become severe if not treated. Moderate reactions include: facial oedema, faintness, severe vomiting, urticaria, bronchospasm

  • Severe: Signs and symptoms are often life-threatening and can result in permanent morbidity or death if not managed appropriately. Severe reactions include: laryngeal oedema, pulmonary oedema, respiratory arrest, hypotensive shock, convulsions, cardiac arrest

  • Delayed: Skin rash, thyrotoxicosis, kidney dysfunction.

The frequency of allergic-like and physiologic adverse events related to the intravascular administration of iodinated contrast media is low and has decreased considerably with changes in usage from ionic, high-osmolality contrast agents (HOCA) to non-ionic, low-osmolality contrast agents (LOCA). Historically, acute adverse events occurred in 5–15 % of all patients who received HOCA. Many patients receiving intravascular HOCA experienced physiologic disturbances (e.g., generalized warmth, nausea, or emesis), and this was often documented as a contrast reaction. HOCA are now rarely or never used for intravascular purposes, because of their greater adverse event profile compared to LOCA. The reported overall acute adverse reaction rate (allergic-like + physiologic) for non-ionic LOCA (i.e., iohexol, iopromide, or iodixanol) ranges from 0.2 %, 0.6 % to 0.7 % [5961]. Serious acute reactions to iv. LOCA are rare, with a historical rate of approximately four in 10,000 (0.04 %) [62].

Reconstruction methods

Myocardial perfusion images are the result of a complex reconstruction process. Although sophisticated reconstruction methods are available, including correction for motion, attenuation and scatter correction, these software tools cannot produce “miracles”. It is therefore important to achieve optimal quality of the raw data by selecting the proper matrix size, angular sampling, zoom factor, patient-to-camera distance, energy window settings, and assuring that the camera is properly tuned and maintained through regular quality control procedures. In addition, the acquired projection data should be checked for motion and the presence of high extra-cardiac uptake. This should be done before the patient leaves the department and before reconstruction is commenced.

Camera vendors and various third party companies provide reconstruction software that implement iterative reconstruction based on Ordered Subset Expectation Maximisation or maximum likelihood expectation maximization (OS-EM/ML-EM). The advantage of these algorithms over traditional filtered-back projection is that information about the camera, patient and radiopharmaceutical can be exploited to reconstruct better images. CT images can be incorporated for estimation of attenuation and scatter; the collimator-detector-response can be modelled and used for resolution recovery; noise can be compensated by modelling the underlying characteristics of the decay process.

These reconstruction methods can achieve enhanced image quality that may be traded against shorter acquisition times or reduced administered activity. Fundamental to all these algorithms is the correct choice of user-selectable parameters (typically number of iterations and subsets, regularisation, and filter parameters). Inadequate parameters most likely lead to insufficient image quality and artefacts. As implementations vary considerably across vendors, it is not possible to transfer settings between camera systems without prior validation.

Currently, all major vendors offer the possibility to include resolution recovery (also called count recovery) in the OS-EM/ML-EM algorithm. The increased reconstructed resolution and lower noise allow for slightly lower count statistics (hence lower injected activity) or shorter scan times [35]. However, such techniques require careful testing against phantom studies performed with validated hardware and software.

Data analysis of hybrid imaging

Hybrid imaging is defined as the combination and fusion of two data sets by which both modalities significantly contribute to image information [63]. Typically, hybrid imaging is synergistic, i.e. more powerful than the sum of its parts, as it provides information beyond that achievable with either data set alone, leading to improved sensitivity and specificity (Fig. 3).

Fig. 3
figure 3

Coronary artery territories in a 17-segment model Myocardial perfusion SPECT, coronary computed tomography angiography (CCTA), and fused hybrid SPECT/CCTA of a 43-year-old male patient with presenting symptoms of typical angina. Myocardial perfusion SPECT documents a reversible perfusion defect in short axis and horizontal long axis slices (a) at rest (bottom rows) and stress (top rows). The corresponding polar plots (b) at rest (left plot) and stress (right plot) clearly depict the extent of the ischaemic area in the anterolateral wall. CCTA (c) shows an intermediate stenosis (i.e., 50–70 % luminal narrowing) due to non-calcified plaques in the middle/distal left anterior descending artery at the level of the second diagonal branch bifurcation. Fused hybrid SPECT/CCTA (d) reveals that the anterolateral ischaemia corresponds with the vascular territory of the second diagonal branch, while the stenosis in the left anterior descending artery (LAD) does not cause any ischaemia

The term hybrid imaging is not valid for the combination of nuclear cardiac imaging with X-ray-based AC where the (low-dose) CT images do not provide independent information, but only contributes to image quality improvement of the other modality (SPECT) [4].

Hardware and software requirements

Dedicated workstations capable of two-dimensional and three-dimensional displays of CCTA and SPECT data are a basic requirement for hybrid data analysis. Aside from projection of standard views of SPECT data, the combination of hardware and software has to offer capabilities for volume rendering of stress and rest SPECT data sets (uncorrected and attenuation corrected), common reconstruction and reformatting techniques for CCTA data [including transaxial stacks, multiplanar (curved) reformations, maximum intensity projections and volume-rendering] [64], and automatic or manual image co-registration and fusion. If non-contrast enhanced CT data is acquired for AC of SPECT, the software should offer computation of coronary artery calcium scores.

Integration of myocardial perfusion SPECT with CCTA

The incremental value of hybrid cardiac imaging arises from spatial co-localization of a myocardial perfusion defect with a subtending coronary artery. Traditionally, this process is performed by mental integration of a standard myocardial segmentation model that allocates each segment to one of the three main coronary arteries [65].

Notably, however, there is a substantial inter-individual anatomical variability of coronary arteries. Therefore, the so-called standard distribution of myocardial perfusion territories does not correspond with the patients’ individual anatomy in more than half of the patients [66]. Most frequently, the left circumflex artery segments are erroneously assigned to the right coronary artery territory, and standard left anterior descending artery segments are erroneously assigned to the left circumflex territory [67]. True hybrid cardiac imaging and data analysis using volume rendered, co-registered, and fused SPECT and CCTA data sets should be preferred over sole side-by-side analysis because of accurate segmental assignment to coronary artery territories and documented incremental value [6870].

Fused images of myocardial perfusion SPECT and CCTA can be obtained using data sets acquired on a hybrid scanner (SPECT/CT) or via co-registration of images obtained from separate stand-alone scanners. Software-based automated registration is accurate and fast [70]. However, automated co-registration of data from cardiac imaging is more challenging than that from other body areas. Errors may arise from cardiac motion, breathing motion, and cardiac contraction [71]. Additionally, different anatomic features are depicted by the two modalities rendering automatic object recognition and image registration difficult, particularly if large perfusion defects are present in SPECT images. Accurate registration, however, is of utmost importance to accurately allocate subtending coronary arteries to areas with radiotracer uptake. Therefore, review of the registration is mandatory and manual correction must be performed if needed.

Data analysis

The combined diagnostic information from myocardial perfusion SPECT imaging and CCTA is complementary: the negative predictive value of CCTA to exclude CAD is close to perfect (Fig. 3) [49, 72, 73]. However, specificity and positive predictive value of CCTA for identification of haemodynamically significant luminal narrowing have been documented to be less robust. CCTA is associated with a general overestimation of the severity of coronary artery stenosis and difficulties to differentiate between slight differences in stenosis severity. SPECT offers complementary diagnostic information as to the haemodynamic significance of coronary artery lesions detected by CCTA. Also, uninterpretable segments on CCTA due to strong calcifications or artefacts can be well studied with SPECT. By contrast, CCTA adds to the diagnostic value of SPECT through documentation of multivessel disease (with possible balanced ischaemia not detectable by semi-quantitative SPECT analysis) or diagnosis of subclinical atherosclerosis.

Interpretation of CCTA studies is beyond the scope of this document and is covered in detail in the respective guidelines of the Society of Cardiovascular Computed Tomography [64]. In brief, CCTA should be first reviewed in transaxial images, complemented by multiplanar reformations to better visualize suspected lesions. Diagnostic conclusions are not based on three-dimensional, volume-rendered CCTA representations.

By contrast, in hybrid imaging, a panoramic three-dimensional view is offered by integrating volume-rendered CT data with the perfusion information from SPECT. This comprehensive information improves both the identification of the culprit vessel and the diagnostic confidence for categorizing intermediate lesions and equivocal perfusion defects, thus optimizing management decisions.

Incidental cardiac and extracardiac CT findings are not uncommon [74]. Although most such findings are negligible, some may occasionally be of clinical relevance. Thus, it is recommended that images are additionally screened by a physician fully trained in CT readings, including non-contrast enhanced CT scans for attenuation correction and scouts.

On a practical level, this concept requires careful coordination of the readout between nuclear physicians and radiologists, and the final discussion with the cardiologist. The benefit of such an interdisciplinary approach is that those fully trained in the specific modalities would interpret the images jointly, thus providing a high-quality result covering all aspects of hybrid SPECT/CCTA image acquisition.

The integration of coronary anatomy by CCTA and (quantitative) documentation of the ischaemic burden through SPECT allows effective identification of: 1) patients with CAD benefitting from optimal medical therapy versus those who should undergo coronary revascularization, 2) the culprit stenosis in patients with multiple coronary artery lesions, thereby guiding clinicians on the appropriate method of revascularization, 3) patients with subclinical coronary atherosclerosis where more aggressive prevention may be indicated, and 4) patients with normal coronary arteries who can safely be deferred from any further cardiac testing.

In summary, the complimentary anatomic and functional information provided by hybrid SPECT/CCTA imaging has been demonstrated to confer added diagnostic value in CAD detection [67, 69, 70, 75, 76] and to effectively stratify risk and predict outcomes [77, 78], guide patient management [79], and to contribute to optimal downstream resource utilization [80].