Myocardial perfusion quantitation with 15O-labelled water PET: high reproducibility of the new cardiac analysis software (Carimas™)

  • Original Article
  • Published:
European Journal of Nuclear Medicine and Molecular Imaging Aims and scope Submit manuscript

An Erratum to this article was published on 27 January 2010

Abstract

Purpose

Carimas™ (Cardiac Image Analysis System) is a new software package developed at the Turku PET Centre for the quantitation of PET studies of the heart with a broad range of tracers. The goal of this study was to assess the reproducibility of results the package provides for myocardial perfusion (MP) quantitation using 15O-labelled water.

Methods

Four observers with various levels of experience in nuclear medicine independently analysed 20 MP studies (10 rest flow: “rest”, 10 adenosine-induced hyperaemia: “stress”). Each study was analysed twice. The linear mixed model for repeated measures was fitted to the data to calculate intraclass correlation coefficients (ICC), differences between the repeats (the intraobserver differences) and differences between the observers (the interobserver differences). Also, Pearson correlation coefficients (r) were calculated and Bland-Altman plots were drawn. The reproducibility of MP was assessed on global, regional and segmental levels. Thereafter, this analysis was applied in 48 consecutive clinical patients with suspected coronary heart disease (CHD).

Results

For the experienced observer the Pearson r for all segments was 0.974 at rest and 0.978 at stress (p < 0.0001), and the repeatability coefficients were 0.145 ml/g per min (15.5% of the average) and 0.389 ml/g per min (14.9%), correspondingly. The ICC reflected very good overall reproducibility. The intraobserver and interobserver differences were small, and the difference between the most and the least experienced observers at stress was 8.5% for the global MP. The clinical accuracy of the perfusion in the detection of CHD was excellent (positive predictive value 91% and negative predictive value 88%) against invasive angiography.

Conclusion

The results demonstrate high reproducibility of myocardial perfusion quantitation with 15O-labelled water PET using Carimas™. The results support the feasibility of robust analysis and good clinical accuracy.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Hartiala J, Knuuti J. Imaging of the heart by MRI and PET. Ann Med 1995;27 1:35–45.

    Article  CAS  PubMed  Google Scholar 

  2. Merhige ME, Breen WJ, Shelton V, Houston T, D’Arcy BJ, Perna AF. Impact of myocardial perfusion imaging with PET and (82)Rb on downstream invasive procedure utilization, costs, and outcomes in coronary disease management. J Nucl Med 2007;48 7:1069–76.

    Article  PubMed  Google Scholar 

  3. Smith MF. Advances in rubidium PET and integrated imaging with CT angiography. Curr Cardiol Rep 2008;10 2:135–41.

    Article  PubMed  Google Scholar 

  4. Wijns W, Camici PG. The value of quantitative myocardial perfusion imaging with positron emission tomography in coronary artery disease. Herz 1997;22 2:87–95.

    Article  CAS  PubMed  Google Scholar 

  5. Pandit-Taskar N, Grewal RK, Strauss HW. Cardiovascular system. In: Christian PE, Waterstram-Rich KM, editors. Nuclear medicine and PET/CT technology and techniques. St. Louis: Mosby Elsevier; 2007. p. 479–512.

    Google Scholar 

  6. Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation 2002;105 4:539–42.

    Article  PubMed  Google Scholar 

  7. Kaufmann PA, Gnecchi-Ruscone T, Yap JT, Rimoldi O, Camici PG. Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET. J Nucl Med 1999;40:1848–56.

    CAS  PubMed  Google Scholar 

  8. Lodge MA, Bengel FM. Methodology for quantifying absolute myocardial perfusion with PET and SPECT. Curr Cardiol Rep 2007;9 2:121–8.

    Article  PubMed  Google Scholar 

  9. Pitkänen OP, Nuutila P, Raitakari OT, Porkka K, Iida H, Nuotio I, et al. Coronary flow reserve in young men with familial combined hyperlipidemia. Circulation 1999;99 13:1678–84.

    PubMed  Google Scholar 

  10. Kaufmann PA, Camici PG. Myocardial blood flow measurement by PET: technical aspects and clinical applications. J Nucl Med 2005;46 1:75–88.

    PubMed  Google Scholar 

  11. Iida H, Rhodes CG, de Silva R, Yamamoto Y, Araujo LI, Maseri A, et al. Myocardial tissue fraction–correction for partial volume effects and measure of tissue viability. J Nucl Med 1991;32 11:2169–75.

    CAS  PubMed  Google Scholar 

  12. Iida H, Rhodes CG, de Silva R, Araujo LI, Bloomfield PM, Lammertsma AA, et al. Use of the left ventricular time-activity curve as a noninvasive input function in dynamic oxygen-15-water positron emission tomography. J Nucl Med 1992;33 9:1669–77.

    CAS  PubMed  Google Scholar 

  13. Cerqueira MD, Verani MS, Schwaiger M, Heo J, Iskandrian AS. Safety profile of adenosine stress perfusion imaging: results from the Adenoscan Multicenter Trial Registry. J Am Coll Cardiol 1994;23:384–9.

    Article  CAS  PubMed  Google Scholar 

  14. Wackers FJT, Bruni W, Zaret BL. Acquisition, processing, display, and analysis of PET images, Chap. 11. Nuclear cardiology: the basics: how to set up and maintain a laboratory. Totowa: Humana; 2008. p. 173–210.

    Google Scholar 

  15. Davis CR. Statistical methods for the analysis of repeated measurements. New York: Springer; 2002. p. 125–55.

    Google Scholar 

  16. Fleiss JL. Reliability of measurements. In: Fleiss JL, editor. The design and analysis of clinical experiments. New York: John Wiley & Sons; 1986. p. 1–32.

    Google Scholar 

  17. Katoh C, Kubo N, Shiga T, Kubo N, Nakada K, Tamaki N. Improvement of algorithm for quantification of regional myocardial blood flow using 15O-water with PET. J Nucl Med 2004;45 11:1908–16.

    PubMed  Google Scholar 

  18. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1 8476:307–10.

    CAS  PubMed  Google Scholar 

Download references

Acknowledgements

We thank Stephen Hitchens (University of Sheffield, Sheffield, UK) for his help with English.

The study was conducted within the “Centre of Excellence in Molecular Imaging in Cardiovascular and Metabolic Research" supported by the Academy of Finland, University of Turku, Turku University Hospital and Åbo Academy.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Sergey V. Nesterov.

Additional information

An erratum to this article can be found at http://dx.doi.org/10.1007/s00259-009-1373-9

Rights and permissions

Reprints and permissions

About this article

Cite this article

Nesterov, S.V., Han, C., Mäki, M. et al. Myocardial perfusion quantitation with 15O-labelled water PET: high reproducibility of the new cardiac analysis software (Carimas™). Eur J Nucl Med Mol Imaging 36, 1594–1602 (2009). https://doi.org/10.1007/s00259-009-1143-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00259-009-1143-8

Keywords

Navigation