Thyroid Probe Counts of 11.1-MBq Capsule of 123I Without Attenuator, Within Neck Phantom, and Behind Cadaver Chest Wall
Object | Data |
---|---|
Background | 115 counts/min |
Capsule | 1.581 × 106 counts/min |
Capsule + neck phantom | 1.290 × 106 counts/min |
Capsule + chest wall | 1.029 × 106 counts/min |
Calculated attenuation | |
Neck phantom | 18% |
Chest wall | 35% |