TABLE 1

Comparison of Acquisition and Processing Parameters Between FASpecT and Standard SPECT for Blood Pool Imaging

ParameterFASpecT/CT protocolStandard SPECT/CT protocol
SPECT acquisition30° steps and total angular range of 360°with dual-head SPECT/CT camera3°–6° steps and total angular range of 360° with dual-head SPECT/CT camera
Acquisition time per projection30 s20–30 s
Total imaging time required3 min 20 s22–65 min
Number of steps660
Total images acquired1260–120
Matrix size128 × 12864 × 64 or 128 × 128
ReconstructionOSEM (2 iterations,10 subsets)Filtered backprojection or OSEM
Postreconstruction filterButterworth: critical frequency, 0.48; power, 10Butterworth with varying parameters