TABLE 1

Technical Parameters of PEM Camera (5)

ParameterCharacteristic
Acquisition
 ViewsAnterior and posterior images of hands and feet
 Field of view  (x–y plane)24 × 16.8 cm (maximum)
 Field of view  (z direction)Patient-dependent (maximum up to 19 cm)
 Compression force67 newton for breast view, patient-dependent for hand and foot view
 Scan durationVariable (typically 10 min)
Reconstruction
 Coincidence timing window6 ns
 Energy window350–750 keV
 Acceptance angle25 crystals; angle varies with paddle separation
 AlgorithmIterative 3-dimensional maximum-likelihood expectation maximization
 Number of iterations5
 CorrectionsDetector normalization and geometric efficiency; no corrections for randoms, dead-time, attenuation, scatter, or intrascan decay
Reconstruction timeDepends on number of counts (typically <15 min)
Images
 Image matrix136 × 200
 Pixel size1.2 × 1.2 mm
 Resolution2.4 mm in full width at half maximum
 Number of slices12
 Slice thickness1/12 detector separation
 UnitsμCi/cm3 or PEM uptake value