TABLE 3

Mean, SD, and P value for 18F-FDG Dose, Scan Time, and DLP for Gemini Vs. Vereos PET/CT Scanner Regarding Different FOVs

18F-FDG dose (MBq)Scan time (s)DLP (mGy/cm)
FOVGeminiVereosPGeminiVereosPGeminiVereosP
Skull to mid thigh443.26 (91.95)370.00 (46.46)<0.0011,342.24 (510)836.40 (139.2)<0.001463.33 (162.14)424.38 (115)NS
Whole body411.99 (95.30)355.94 (62.24)<0.0011,619.95 (595.8)1,083.24 (180.6)<0.001702.22 (261.36)567.46 (144.28)<0.001
  • NS = not significant.

  • Data are mean followed by SD in parentheses. P values were determined by t test.