TABLE 1

Optimal Reconstruction Parameters for 89Zr Oncology Imaging

Reconstruction parameters
ScannerAlgorithmIterationsSubsetsGaussian (mm)PSF
DSTOSEM2214Not available
600OSEM, 3D3167SharpIR
690OSEM3167SharpIR
710VPFX2247SharpIR
mCTTrue TOF3217Not used
  • PSF = point-spread function; OSEM = ordered-subset expectation maximization; 3D = 3-dimensional; SharpIR = point-spread function correction; VPFX = GE Healthcare standard reconstruction algorithm; TOF = time of flight.