TABLE 1

Recommended Quality Control Testing Program for MBI Systems

TestEquipmentFrequencyAcquisition detailsPassing criterion
Uniformity57Co sheet source  or fillable phantomDaily7.5 million counts≤5% integral uniformity
Spatial resolution4-quadrant bar phantomSemiannually7.5 million counts; phantom  angled across field of viewMeets manufacturer’s  specifications
SensitivityFlaskAnnually120-s images≤10% difference between  2 detectors
Energy resolutionPoint sourceAnnually2-keV energy windows; 1-min  imagesFull width at half maximum  ≤ 10%
Lesion contrastContrast-detail phantomQuarterly1 million counts; images at  3 depthsCNR > 3; count number of  visible lesions at each depth
  • * All tests should be performed at acceptance testing and after major service work.

  • CNR = contrast-to-noise ratio.