TABLE 1

RCP Evaluation Outcomes with Various Mobile Phases

ITLC-SG paper
Mobile phaseImpurity Rf68Ga-DOTATATE RfDevelopment time (min) (n = 3)Rs
Methanol:1 M ammonium acetate
 50:500.0–0.10.8–1.025.3 ± 1.72.3
 60:400.0–0.10.8–1.023.9 ± 0.52.2
 80:200.0–0.10.6–0.918.6 ± 0.42.1
 90:100.0–0.10.3–0.615.1 ± 0.30.6
 95:50.0–0.10.1–0.313.9 ± 0.3NA
100:00.0–0.10.0–0.110.3 ± 0.2NA
 0:1000.0–0.10.0–0.110.3 ± 0.6NA
  • NA = unable to identify 2 peaks on ITLC-SG.