TABLE 2

Noise Comparison for Reconstructions with MLEM and Windowed FBP, When Images Are at Same Lesion Contrast

Iteration no. for MLEMk for windowed FBPNormalized noise SD for MLEM imageNormalized noise SD for windowed FBP image
53,8000.02860.0277
108,2000.05440.0530
1512,0000.07860.0711
2017,0000.10150.0904
2522,0000.12330.1058
3028,0000.14420.1205
3533,0000.16430.1303
4038,0000.18370.1384