TABLE 1

Results for Each of the Initial 40 Patients

Physician reportingSensor reporting
Patient no.Infiltration present?Would SUV be affected?Infiltration present?Additional characterization?
1NoYes, minorYes
2NoNoNo
3Site not in FOVYes, moderateYes
4NoNoNo
5Yes, minorNot likelyYes, minorNo
6NoYes, minorYes
7Yes, minorNot likelyYes, moderateYes
8NoNoNo
9Site not in FOVYes, minorYes
10Yes, moderateNot likelyYes, moderateNo
11NoNoNo
12Site not in FOVNoYes
13Yes, moderateNot likelyYes, minorYes
14Yes, minorNot likelyYes, minorNo
15Yes, significantVery likelyYes, significantNo
16NoNoNo
17Yes, moderateNot likelyYes, moderateNo
18Site not in FOVNoYes
19Yes, moderatePossibly likelyYes, moderateNo
20Yes, minorNot likelyYes, minorNo
21Yes, minorNot likelyYes, minorNo
22NoNoNo
23NoNoNo
24NoNoNo
25NoNoNo
26*NoYes, minorYes
27Yes, moderateNot likelyYes, minorYes
28NoYes, minorYes
29NoNoNo
30NoNoNo
31YesNot likelyYes, minorNo
32NoNoNo
33Yes, minorNot likelyYes, minorNo
34Yes, minorNot likelyYes, minorNo
35Yes, minorNot likelyYes, minorNo
36NoNoNo
37NoNoNo
38NoNoNo
39Yes, minorYes, minorYes
40NoNoNo
  • * Injection was in right forearm and within FOV; no injection problems were evident.

  • FOV = field of view.