TABLE 2

Physicochemical Results for Controls

pHRadiochemical purity (%)
AgentConformityBaseline4 or 7 h*ConformityBaseline4 or 7 h*
99mTc-pertechnetate eluate (n = 5)4–85.84 (5.8–6.0)5.80 (5.8–5.8)>95%99.63 (99.37–99.99)99.56 (99.34–99.99)
99mTc-HMDP (n = 5)5–75.36 (5.0–5.5)5.36 (5.0–5.5)>95%99.63 (99.34–99.79)99.05 (98.42–99.57)
99mTc-HSA (n = 5)2–6.55.4 (5.0–6.0)5.4 (5.0–6.0)>95%99.70 (99.05–99.99)99.51 (99.13–99.76)
99mTc-albumin nanocolloid (n = 5)4–75.0 (5.0–5.0)5.0 (5.0–5.0)>95%99.55 (99.31–99.99)99.67 (99.10–99.99)
  • * 7 h for 99mTc-pertechnetate eluate, 99mTc-HMDP, and 99mTc-HSA, and 4 h for 99mTc-albumin nanocolloid.

  • Data are mean with or without range in parentheses.