TABLE 2

Comparison of Accuracy of Administered 18F-FDG Radioactivity Between Automated Infusion Device and External Dose Calibrator

DevicesSampleRadioactive concentration (MBq/mL)Dispensed volume of 18F-FDG (mL)Radioactivity recorded by device (MBq)External measurement (MBq)Administration accuracy (%)Radioactive concentration (MBq/mL)Dispensed volume of 18F-FDG (mL)Radioactivity recorded by devise (MBq)External measurement (MBq)Administration accuracy (%)
M-1301294.120.1748.4250.33−3.79225.940.84189.79192.89−1.60
2121.950.4148.4049.76−2.72127.321.40178.25179.58−0.74
3200.000.2549.2650.05−1.59202.310.87176.01178.43−1.36
4147.060.3448.5249.84−2.66152.301.17178.20180.36−1.20
5166.670.3048.1949.97−3.56167.531.04174.23172.640.92
6142.860.3524.0724.42−1.4259.611.5491.8094.27−2.63
AI-3001104.170.4848.8747.872.09118.801.42168.70169.63−0.55
2131.580.3850.0949.341.52165.571.12185.44170.628.68
3147.060.3450.6144.5413.64178.591.04185.73175.665.73
4131.580.3852.0146.8810.94161.711.14184.35173.836.05
5151.520.3349.0142.8314.42177.281.05186.14171.398.61
6161.290.3148.9347.981.99183.621.01185.46182.751.48
UG-05190.790.6054.4851.276.2692.042.01180.23174.923.04
2154.730.3452.6150.164.87181.371.02147.90141.924.21
3205.090.2653.3252.142.27205.560.90185.01175.615.35
4206.850.2653.7850.566.37179.611.03179.35177.291.16
5224.450.2556.1152.985.92212.640.87182.29173.105.31
6108.580.3942.3442.40−0.13411.110.45191.05187.631.83