TABLE 1

Comparison of Image Quality at Acceptance Testing and 12 Months Later

Insert diameter
Parameter10 mm13 mm17 mm22 mm28 mm37 mm
Contrast recovery
 At acceptance testing34.70%51.43%60.83%72.13%73.8877.65%
 12 mo later36.30%50.62%60.13%70.69%72.1276.92%
 Difference+1.6%−0.81%−0.60%−1.44%−1.76%−0.73%
Background variability
 At acceptance testing8.65%7.89%7.05%6.15%5.46%5.04%
 12 mo later8.96%8.01%6.85%6.65%5.77%5.32%
 Difference+0.31%+0.12−0.20%+0.50%+0.31%+0.28%
  • Scatter correction using filtered backprojection reconstruction was applied.