Detailed Protocol for BET
Test serial no. | Assay tube (in duplicate) | LAL reagent water (μL) | Control standard endotoxin 4λ (μL) | Test sample* (μL) | LAL reagent (μL) |
1 | Negative water control | 100 | — | — | 100 |
2 | Positive water control 2λ | 50 | 50 | — | 100 |
3 | NPC† | 50 | — | 50 | 100 |
4 | PPC 2λ† | — | 50 | 50 | 100 |
5 | PPC 2λ with calcium chloride† | — | 50 | 50‡ | 100 |
↵* Repeated at MVD and MVD/2—that is, 1:66 and 1:33 dilutions.
↵† These tubes were set up for all 99mTc radiopharmaceuticals (99mTc-DTPA, 99mTc-DMSA-III, 99mTc-DMSA-V, 99mTc-ECD, 99mTc-mebrofenin, 99mTc-methoxyisobutylisonitrile, and 99mTc-MDP) and their cold kits.
↵‡ Calcium chloride at final concentration of 300 μM was incorporated in assay tube of test sample at final dilution.
Tests 3 and 4 were also performed initially for 60 mM stock solution of calcium chloride and saline (0.9% sodium chloride) to ensure absence of endotoxin in stock solutions.