Results for Different 99mTc-Labeled Radiopharmaceuticals and Their Cold Kits at MVD and MVD/2 in BET Assay
MVD | MVD/2 | |||||||
NPC | PPC | NPC | PPC | |||||
Product analyzed | Without CaCl2 | With CaCl2 | Without CaCl2 | With CaCl2 | Without CaCl2 | With CaCl2 | Without CaCl2 | With CaCl2 |
99mTc-EC | – | – | + | ++ | – | – | – | ++ |
EC cold kit | – | – | – | ++ | – | – | – | ++ |
99mTc-DTPA | – | – | + | ++ | – | – | – | ++ |
DTPA cold kit | – | – | – | ++ | – | – | – | ++ |
99mTc-ECD | – | – | ++ | ++ | – | – | ++ | ++ |
ECD cold kit | – | – | – | ++ | – | – | – | ++ |
99mTc-DMSA-III | – | – | – | ++ | – | – | – | ++ |
DMSA-III cold kit | – | – | – | ++ | – | – | – | ++ |
99mTc-DMSA-V | – | – | – | ++ | – | – | – | ++ |
DMSA-V cold kit | – | – | – | ++ | – | – | – | ++ |
99mTc-mebrofenin | – | – | ++ | ND | – | – | ++ | ND |
Mebrofenin cold kit | – | – | + | ++ | – | – | – | ++ |
99mTc-methoxyisobutylisonitrile | – | – | ++ | ND | – | – | ++ | ND |
methoxyisobutylisonitrile cold kit | – | – | ++ | ND | – | – | ++ | ND |
99mTc-MDP | – | – | ++ | ND | – | – | ++ | ND |
MDP cold kit | – | – | ++ | ND | – | – | ++ | ND |
– = no gel-clot; + = semisolid gel-clot; ++ = firm gel-clot; ND = not done.