Pitfalls in Use of a Scintillation Detector
Type of use | Pitfall |
Routine | Geometry of measurement not according to protocol |
Calibration performed incorrectly | |
Incorrect energy window chosen | |
Extraneous radiation source present (patient, contamination, etc.) | |
Background not subtracted | |
Source activity too high; measurement incurs dead time | |
Efficiency factor not applied when needed | |
Incorrect detector chosen (well counter vs. thyroid probe) | |
QC | Daily calibration not performed or performed incorrectly |
Geometry of constancy measurement not according to protocol | |
χ2 test affected by extraneous sources of radiation | |
Efficiency factor out of date or not determined according to protocol | |
Current source activity not calculated correctly for efficiency factor determination | |
Energy window incorrect for given QC test |