TABLE 3

Influence of 99mTc Generator Eluates on RCP Values of Different Radiopharmaceuticals

NCWC generatorNFDC generatorNFWC generator
RadiopharmaceuticalMethodNMean ± SDNMean ± SDNMean ± SDP
99mTc-succimer3199.77 ± 0.281999.82 ± 0.21999.49 ± 0.99NS
99mTc-oxidronate13197.65 ± 0.70NTNT3197.54 ± 0.85NS
23198.09 ± 1.223198.14 ± 0.91NTNTNS
99mTc-tetrofosmin13195.39 ± 1.32NTNT3194.13 ± 0.890.0002
23198.28 ± 0.543198.14 ± 0.88NTNTNS
99mTc-exametazime13190.51 ± 2.25NTNT3189.51 ± 2.280.0532
23191.33 ± 1.923190.39 ± 2.13NTNTNS
99mTc-albumin nanocolloid3299.88 ± 0.142299.84 ± 0.163299.86 ± 0.18NS
99mTc-soluble albumin3597.31 ± 4.383897.89 ± 1.36NTNTNS
  • NT = not tested; NS = not significant.

  • For 99mTc-oxidronate, method 1 used ITLC/SG strips from Varian SA and method 2 used ITLC/SG strips from Pall Corp. For 99mTc-tetrofosmin, method 1 used chromatography paper from Varian SA and method 2 used ITLC/SG strips from Pall Corp. For 99mTc-exametazime, method 1 used chromatography paper from Varian SA and method 2 used ITLC/SG strips from Pall Corp. P values compared data from NCWC generators with those obtained from NFDC or NFWC generators.