TABLE 3

CFs for Measurements Using Copper Sleeves

Dose calibratorVial CFSyringe CF (10 mL)Difference (syringe CF − vial CF)Manufacturer and model
11.288Not doneBiodex Atomlab 100
21.0521.0070.045Capintec CRC-15R
31.6631.6280.035Capintec CRC-15R
42.3952.3600.035Capintec CRC-15R
50.9780.9440.034Capintec CRC-15R
60.9760.9710.005Amersham Radioisotope Calibrator (ARC)-150
72.5302.4610.069Capintec CRC-15R
82.5662.4920.074Capintec CRC-15R
90.9310.8870.044Capintec CRC-12R
101.0041.0700.064Capintec CRC-15R
Mean1.5381.5360.002