TABLE 5

Vial CF by Dose Calibrator Manufacturer: No Copper Sleeve

ManufacturerNo. of unitsMean vial CF
Biodex231.438 (range, 1.064–1.626)
Capintec1181.274 (range, 0.823–1.757)
Empos Ltd.10.988
Medisystem20.922 (range, 0.890–0.954)
MIE Medical Imaging Electronics GmbH11.312
Nuclear Associates11.298
Nuklear Medizintechnik91.193 (range, 0.977–1.534)
PTW Freiburg GmbH61.069 (range, 0.917–1.207)
Radcal11.227
Syncor11.224
Veenstra41.210 (range, 0.996–1.332)