TABLE 2

Correlation Between Size of Rods Without 18F-FDG Activity and Percentage Discerned by VOI Technique

Rods discerned/total rods
Rod dimension (mm)First session (background, 37 kBq [1 μCi]/mL)Second session (background, 111 kBq [3 μCi]/mL)
4.80/57 (0)3/57 (5.3)
6.47/32 (21.8)7/32 (21.8)
7.916/21 (76.8)14/21 (66.6)
9.515/15 (100.0)15/15 (100.0)
11.110/10 (100.0)10/10 (100.0)
12.78/8 (100.0)8/8 (100.0)
  • Spearman unilateral correlation r = 0.984 (P < 0.01). Data in parentheses are percentages.