TABLE 1

Average Exit Dose Rates

Exit dose rate (10−2 μGy/s)*
Voltage (kV)Current (mA)0 mm50 mm100 mm150 mm200 mm250 mm300 mm
1402.52,1381,1005222351104824
1402.01,705887426193873918
1401.51,295675323148672915
1401.088346021810544209
1202.51,396713332147652913
1202.01,126574264117532310
1201.58534172018740177
1201.05802801406128125
  • * Data are for the stated different depths of water in a cylindric phantom simulating attenuation in a patient. Rates are calculated by dividing cumulative exit doses by time of measurement.