TABLE 2

Effect of Labeling Activity and Storage Form on Radiochemical Impurities in Stabilized 99mTc-d,l-HMPAO in Vials and Syringes over Time

1,110 MBq2,600–3,700 MBq
99mTcO2 + secondary 99mTc-HMPAO99mTcO499mTcO2 + secondary 99mTc-HMPAO99mTcO4
Time (h)VialSyringeVialSyringeVialSyringeVialSyringe
04.8 ± 0.9NA2.1 ± 0.9NA6.2 ± 1.2NA2.3 ± 0.7NA
25.5 ± 0.96.2 ± 1.22.4 ± 0.54.0 ± 1.67.0 ± 1.48.1 ± 1.54.1 ± 1.26.0 ± 1.2
46.1 ± 0.86.3 ± 0.92.8 ± 0.65.0 ± 1.78.6 ± 1.79.3 ± 1.96.2 ± 1.710.4 ± 2.5
  • NA = not applicable.

  • Values represent percentage of radiochemical impurities (mean ± SD); n = 10. P < 0.01 for 99mTcO2 + secondary 99mTc-HMPAO and 99mTcO4 between both groups of activities in vial and syringe at all times. P > 0.05 for 99mTcO4 between both groups of labeling activities in vial at t = 0.