TABLE 1

CTDIw vs. Beam Collimation

Total beam collimation (mm)Detector configurationCTDIw relative to 10-mm beam
54 × 1.25 mm1.22
104 × 2.5 mm1.00
154 × 3.75 mm0.93
2016 × 1.25 mm0.89
4016 × 2.5 mm0.83
  • Because discarded penumbra represents smaller fraction of total beam width, CTDI values decrease with increasing beam widths (combined data from GE Healthcare 4-slice and 64-slice MSCT scanners).