RCPs of 99mTc-ATC Determined by ITLC and SAX Methods (n = 3)
RCP (%) | ||||||
---|---|---|---|---|---|---|
30 min | 35 min | |||||
Method | ITLC | SAX | |Δ| | ITLC | SAX | |Δ| |
Water bath at 100°C | 97.6 ± 0.9 | 98.0 ± 0.8 | 0.4 ± 0.1 | 98.1 ± 0.5 | 98.2 ± 0.5 | 0.1 ± 0.0 |
Dry block at 21-mm and 100°C | 96.9 ± 0.4 | 97.3 ± 0.4 | 0.4 ± 0.0 | 97.6 ± 0.3 | 97.9 ± 0.3 | 0.3 ± 0.0 |
Dry block at 21-mm and 115°C | 99.1 ± 0.2 | 99.3 ± 0.1 | 0.2 ± 0.1 | 99.3 ± 0.4 | 99.3 ± 0.3 | 0.2 ± 0.1 |
Dry block at 26-mm and 115°C | 98.0 ± 0.4 | 98.3 ± 0.4 | 0.4 ± 0.1 | 98.5 ± 0.6 | 98.4 ± 0.2 | 0.4 ± 0.1 |