TABLE 1

Acquisition Parameters for Standard Lung Perfusion and Ventilation Imaging in 18 Participating Hospitals

HospitalCamera age (y)CollimatorMatrixPatient positionPerfusion dose (MBq)No. of projectionsFrame (kcts or time)Ventilation methodDose (MBq)Frame (kcts or time)No. of projections
TU8HR512Supine1116300 sVenticol740600 s6
IJ10AP256Supine504300 sVenticol1,100300 s4
B6HR256Supine1856600Venticol111–1,1101004
H10HS256Supine1306600Solcovent2,0003006
E10GP256Supine1486700Technegas22–33300 s4
PQ13HR128Supine1858130 sSolcovent2,220180 s8
PQ210AP128Supine156600+tSolcovent2,220600+t6
V115GP256Supine1106600Technegas20–402006
V210AP128Supine1106600Technegas20–402006
FG16AP128Supine111–1586600Technegas3704006
R8HR128Supine11061,000+tVenticol900400+t6
N12HR128Supine185690 sVenticol740150+t6
Z16AP128Supine1484600Technegas374004
S12AP256Supine1006400Technegas4002006
XY4GP128Supine1006120 sTechnegas300120 s6
M12HS128Supine746400Venticol1,3004006
K21GP256Supine11041,000Aerosol740300 s4
O1AP256Supine150–185890 sTechnegas150–400120 s8
CD11GP256Supine1114400Venticol740150–3004
L15GP256Sitting744600/500Ultravent1,4803004
  • HR = high resolution; AP = all purpose; HS = high sensitivity; GP = general purpose; 600+t = first frame is images at 600 kcts and rest are with same time.

  • Hospitals are coded to maintain anonymity.