Well Counter Reproducibility for Simulated (Flumazenil) HPLC Metabolite Curves (55 Repeated Measurements)
High activity | Low activity | |
---|---|---|
Total activity | Mean = 61,968 cpm | Mean = 1,462 cpm |
SD = 267 cpm (COV = 0.4%) | SD = 73 cpm (COV = 5.0%) | |
Peak 1 | 4 tubes | 5 tubes |
Mean = 10,981 cpm | Mean = 477 cpm | |
SD = 85 cpm (COV = 0.8%) | SD = 25 cpm (COV = 5.2%) | |
Peak 2 | 10 tubes | 7 tubes |
Mean = 4,435 cpm | Mean = 599 cpm | |
SD = 74.9 cpm (COV = 1.7%) | SD = 32 cpm (COV = 5.3%) | |
Peak 3 | 11 tubes | 7 tubes |
Mean = 46,565 cpm | Mean = 381.6 cpm | |
SD = 238 cpm (COV = 0.5%) | SD = 36 cpm (COV = 9.4%) |