TABLE 1

A Protocol for Digital 3-Phase Planar Bone Scanning

Intravenous dose of 99mTc-MDP or -HDPAcquisition phase
AdultChildFlowBlood pool2- to 4-h delayed*
740–1,110 MBq (20–30 mCi)9.25 MBq/kg (250 μCi/kg)Low-energy all-purpose (or high-resolution) collimatorLow-energy all-purpose (or high-resolution) collimatorLow-energy high-resolution collimator
20% energy window centered at 140 keV 128 × 128 matrix 2–5 s/frame for 60 s256 × 256 matrix 60-s images256 × 256 matrix 150 to 1,000 kcts, depending on body part (Table 2) For limited studies on patients with localized pain, acquisition of at least 1 view to include major joints proximal and distal to the painful site
  • * Delay is greater for elderly, debilitated, or diabetic patients.