TABLE 1

Sample Data and Calculations from Patient Dosimetric Study

ParameterDay 0Day 4Day 6
Dose calibrator measurements
 Time of measurement13:2012:5111:30
131I standard activity (kBq [μCi])9,694 (262)6,993 (189)5,846 (158)
Gamma-camera counts (whole-body mode)
131I source
  Time started13:1112:5211:19
  Anterior counts35,92524,30920,411
 Background
  Time started13:0512:4211:14
  Anterior counts2,5442,5102,345
 Patient
  Time started14:1313:0111:27
  Anterior counts327,581125,24874,381
Quality control calculation
 Time from initial count (h)095.7142.1
 Background-corrected source counts33,38121,79918,066
 % initial count1006554
 Counts per 37 kBq (μCi) in standard127.4115.3114.3
Residence time calculation
 Background-corrected patient counts325,037122,73872,036
 % initial count1003822
 Residence time (h) from graph9896
Therapy dose calculation
 Administered 131I dose
  Embedded Image75.476.9
  • Whole-body scan speed, 100 cm/min; collimator, medium energy; patient platelet count, 133,000; patient height, 165 cm; maximum effective mass (from table), 78.7 kg; camera height above table, 36 cm; whole-body scan field of view, 179 cm; desired total-body dose, 65 cGy; patient weight, 82.5 kg; activity hours (from table), 8,523.