Ratio of Glass Vial Measurements to Syringe Measurements
Emitter | VEENSTRA VDC-405 | Atomlab 100 | Isomed 1000/2000 | Capintec CRC | Centronic IG12 | Capintec β |
---|---|---|---|---|---|---|
32P | 0.97 | 0.92 | 1.01 | 0.92 | 1.02 | 1.28 |
89Sr | 0.94 | 0.89 | 0.98 | 0.89 | 1.01 | 1.03 |
90Y | 1.08 | 1.02 | 0.65 | 1.02 | 1.07 | 1.24 |
169Er | 0.77 | 0.73 | 0.86 | 0.69 | 0.96 | 0.96 |