Intrinsic Efficiency of Dose Calibrators for β-Emitters (Rn)
β- Emitter | VEENSTRA VDC-405 | Atomlab 100 | Isomed 1000/2000 | Capintec CRC | Centronic IG12 | Capintec β |
---|---|---|---|---|---|---|
32P | 0.97 | 1.05 | 1.07 | 1.06 | 0.97 | 0.96 |
89Sr | 0.83 | 0.90 | 0.89 | 0.91 | 0.91 | 0.85 |
90Y | 1.59 | 1.73 | 2.80 | 1.71 | 2.01 | 1.47 |
169Er | 0.068 | 0.078 | 0.081 | 0.078 | 0.024 | 0.041 |