TABLE 1

Summary of Metallic Plates/Sheets Used for Cranioplasty and Phantom Study

Type of Metallic Plate/SheetAluminumStainless SteelTantalumLead
Mass attenuation coefficient for 140 keV photon (cm2/g)0.143NA1.825 2.393
Density (g/cm3)2.70NA16.6511.35 
Reported metallic plate thickness used for cranioplasty (cm)0.0640.0460.038, 0.051NA
Theoretical attenuation for reported metallic plate (%)2NA68, 79NA
Experimental metallic sheet thickness used for phantom study (cm)0.0640.061NA0.038, 0.076
Theoretical attenuation for experimental metallic sheet (%)2NANA64, 87
Calculated attenuation from planar image (%)49NANA, 83
  • Mass attenuation coefficient and density information obtained from www.physics.NIST.gov Web site’s reference tables of x-ray mass attenuation coefficients.