TABLE 2

Minimum Detectable Defect Thickness for 99mTc and 201Tl

IsotopeROR (cm)Dmin (mm; percentage of myocardial phantom wall thickness)
ANTSEPIPLATAP
99mTc21.04  (33%)4  (33%)2  (16%)4  (33%)3  (25%)
25.04  (33%)4  (33%)2  (16%)4  (33%)4  (33%)
29.24  (33%)5  (43%)3  (25%)4  (33%)4  (33%)
201Tl21.06  (50%)6  (50%)3  (25%)5  (43%)4  (33%)
25.06  (50%)6  (50%)3  (25%)6  (50%)4  (33%)
29.26  (50%)7  (58%)4  (33%)6  (50%)4  (33%)