Acquisition Parameters: PET or PET/CT
Parameter | Specification | Standard/preferred/optional |
---|---|---|
Camera type | PET/CT | Preferred |
PET | Standard | |
Energy peak | 511 keV | Standard |
Injection-to-imaging time | 60–90 min | Standard |
Attenuation correction | PET/CT: CT acquisition | Standard |
PET: cesium or germanium sources | Standard | |
Patient position | Supine | Standard |
Arm position | Arms overhead | Standard |
Acquisition mode | 2- or 3-dimensional | Standard |
Number of bed positions | Varies | Standard |
Time/bed position | Partial whole-body scan: | Standard |
2-dimensional: 4 min | ||
3-dimensional: 3 min | ||
Dedicated cardiac scan: | ||
2-dimensional: 20–30 min | ||
3-dimensional: 10 min | ||
View | Chest, including the lungs and mediastinum | Standard |
CT acquisition | Per the manufacturer’s recommendations for attenuation correction or diagnostic imaging | Optional |