Acquisition Parameters: SPECT or SPECT/CT
Parameter | Characteristic | Standard/preferred/optional |
---|---|---|
Camera type | Large field of view | Standard |
Energy peak | 159 keV | Standard |
Energy window | 20% | Standard |
Collimator | Low-energy high resolution | Standard |
Patient position | Supine | Standard |
Camera position | Anterior | Standard |
Injection to imaging time | Early: 15 min | Standard |
Late: 4 h | ||
Acquisition type | SPECT | Standard |
Views | NA | NA |
Additional views | NA | NA |
Orbit | 180° | Standard |
Orbit type | Circular | Standard |
Noncircular | Optional | |
Pixel size | 6.4 mm | Standard |
Matrix | 64 × 64 | Standard |
128 × 128 | Optional | |
Number of projections | Single head 60 (6°) | Standard |
Dual head 60/head (3°) | ||
Time/projection | 30 s | Standard |
CT Acquisition | Per manufacturer’s recommendations for attenuation correction or diagnostic imaging | Optional |
NA = not applicable.