TABLE 3.

123I-MIBG Scintigraphy Imaging Parameters

ParameterCharacteristicStandard/preferred/optional
Camera typeLarge–field-of-view γ-cameraStandard
Energy peak159 keVStandard
Energy window20%Standard
CollimatorLow-energy, high resolution, parallel holeStandard
Patient positionSupineStandard
Field of viewHeart/chestStandard
Injection-to-imaging time (delayed images)Late images at 3 h 50 minStandard
PlanarStandard
 Acquisition typeStaticStandard
 Detector configuration180°Standard
 ViewsAnterior and lateralStandard
 Number of views2Standard
 Counts/time per view10 min per imageStandard
 Matrix256 × 256 (minimum 128 × 128)Standard
 MagnificationZoom = 1Standard
SPECT or SPECT/CTOptional
 Acquisition typeStep and shootOptional
 Patient positionSupineOptional
 Orbit360°Optional
 Matrix128 × 128Optional
 MagnificationZoom = 1Optional
 Pixel sizePer camera settingsOptional
 Projections per detector120 steps (60 steps/detector)Optional
 Time per projection30 s per stepOptional
CT attenuation correctionStandard attenuation correctionOptional
CT parameters120 kV, 25 mA, pitch = 1.25Optional
CT parametersSlice thickness, 2.5 mm; 1 s/rotationOptional