TABLE 3.

Acquisition Parameters: SPECT or SPECT/CT

ParameterSpecificationStandard/preferred/optional
Camera typeLarge-field-of-view multidetectorStandard
Energy peak140 keVStandard
Energy window20%Standard
CollimatorLow-energy, high-resolutionStandard
Patient positionSupineStandard
Camera positionLungs within field of viewStandard
Inhalation-to-imaging timeAcquisition started on completion of inhalationStandard
Acquisition typeStep and shoot or continuousStandard
Orbit180° or 360°Standard
Orbit typeNoncircularStandard
Matrix128 × 128Standard
Number of projections120Standard
Time per projection12–15 sStandard
CT acquisitionPer manufacturer’s recommendations for attenuation correction or diagnostic imagingOptional