Acquisition Parameters: SPECT or SPECT/CT
Parameter | Specification | Standard/preferred/optional |
---|---|---|
Camera type | Large-field-of-view multidetector | Standard |
Energy peak | 140 keV | Standard |
Energy window | 20% | Standard |
Collimator | Low-energy, high-resolution | Standard |
Patient position | Supine | Standard |
Camera position | Lungs within field of view | Standard |
Inhalation-to-imaging time | Acquisition started on completion of inhalation | Standard |
Acquisition type | Step and shoot or continuous | Standard |
Orbit | 180° or 360° | Standard |
Orbit type | Noncircular | Standard |
Matrix | 128 × 128 | Standard |
Number of projections | 120 | Standard |
Time per projection | 12–15 s | Standard |
CT acquisition | Per manufacturer’s recommendations for attenuation correction or diagnostic imaging | Optional |