Acquisition Parameters: PET or PET/CT
Parameter | Specification | Standard/preferred/optional |
---|---|---|
Camera type | PET/CT | Preferred |
PET | Standard | |
Energy peak | 511 keV | Standard |
Injection to imaging time | ||
18F-florbetapir | 30–50 min after injection | Standard |
18F-florbetaben | 45–130 min after injection | Standard |
18F-flutemetamol | 60–120 min after injection | Standard |
Attenuation correction | PET/CT: CT acquisition | Standard |
PET: cesium or germanium sources | Standard | |
Patient position | Supine | Standard |
Arm position | Arms down | Standard |
Acquisition mode | 2- or 3-dimensional | Standard (camera-dependent) |
Bed positions | 1 | Standard |
Time per bed position | ||
18F-florbetapir | 10 min | Standard |
18F-florbetaben | 15–20 min | Standard |
18F-flutemetamol | 10–20 min | Standard |
View | Head | Standard |
CT acquisition | Per the manufacturer’s recommendations for attenuation correction or diagnostic imaging | Optional |