Camera type | Large-field-of-view multidetector | Optional |
Energy peak | 140 keV | Optional |
Energy window | 20% | Optional |
Collimator | Low energy, high resolution | Optional |
Patient position | Supine | Optional |
Camera position | To ensure field of view includes kidneys | Optional |
Injection-to-imaging time | 2–4 h after injection | Optional |
Acquisition type | Step and shoot or continuous | Optional |
Orbit | 180° | Optional |
Orbit type | Circular or noncircular | Optional |
Matrix | 128 × 128 | Optional |
Number of projections | 120 | Optional |
Time per projection | 25–30 s | Optional |
CT acquisition | Per manufacturer’s recommendations for attenuation correction or diagnostic imaging | Optional |